Microscopic Examination of Urine Flashcards
The main purpose is to identify and/or quantify
insoluble materials in the urine such as red blood cells (RBs), white blood cells (WBCs), epithelial cells, casts, bacteria, yeast, parasites, mucus, spermatozoa, crystals, and artifacts
MICROSCOPIC EXAMINATION
MICROSCOPIC EXAMINATION is where we detect and identify __________ materials.
INSOLUBLE
In ROUTINE ANALYSIS, _______ specimen must be utilized
FRESH
The refrigeration of urine can lead to the precipitation of _____________ (alkaline pH) and ______________ (acidic pH)
Amorphous phosphates; Amorphous urates
In the three step analysis, what should be done in order in urine testing?
Gross Appearance, Biochemistry, Microscopy
Before performing microscopic
examination is that the specimen must be ________________.
Centrifugation
Urine is prepared in ________ tube (conical vial), before microscopic examination?
Falcon Tube
The urine sample must be centrifuge at ______ rpm for ______ minutes at ______ temperature, to avoid lysis of the formed elements
1,800; 10; 15-30 degree celsius
T/F: In the Philippines, we are using Decislide as we are also using conventional microscopy.
False- In the Philippines, we are still not using
Decislide but we are still using conventional
microscopy
Identify the magnification of microscope:
Scanning: _____
LPO: _______
HPO: ______
4x ; 10x; 40x
The total magnification is evaluated by:
Magnification X Total Eyepiece
In determining cast counts at the microscope, what magnification should be used?
10x
In determining cellular counts, what magnification should be used?
40x
T/F: The condenser must be down at a certain level to have proper contrast of hyaline cast?
TRUE
T/F: HYALINE CAST can be mistaken with Squamous Epithelial Cells.
TRUE
If a formed element is approximated under LPO, but is mistakenly read and graded under HPO, then you’ll tend to have false ________ results; If a cellular element is supposed to be read under
HPO, but is mistakenly read under LPO, then there will
be a false ______ in results.
Decrease; Increase
T/F: In common laboratory practice, when using 10ml/5mL test tube, turning it over and decanting is usually done.
TRUE
Where does urine sediment gets trapped at the top or bottom?
BOTTOM
T/F: After centrifugation, urine sediments are discarded before examining microscopically?
FALSE: Do not discard the sediments. Make sure that you have adequate sediments to be examined microscopically.
When does urine reagent strip testing must be done? Is it BEFORE or AFTER centriugation?
BEFORE - Urine reagent strip testing is done in a homogenous specimen and not after centrifugation.
Performing urine reagent strip testing after centrifugation can cause _____.
FALSE DECREASE RESULT
Is there a standardized amount of resuspended urine
sample that needs to be dispensed?
There is no standardized amount.
Is there a difference in reading when you put
coverslip in the reading?
Yes, reading of cellular elements tends to be HIGHER if you do not use coverslip since your reading is not standardized.
T/F: The importance of using a coverslip is it distribute the cellular elements.
TRUE