Microscope Parts & Functions / Koehler Illumination Flashcards
The enhancement on the Imaging system will work less with what?
Dark stained chromosomes
What are the the Kohler Illumination steps?
1-Center and focus the bulb.
2-Focus on the specimen with low power objective.
3-Close the field diaphragm completely.
4-Rack the condenser up or down until the field diaphragm edges are in focus.
5-Center the field diaphragm image.
6-Open the field diaphragm until the edges disappear.
7-Set the condenser diaphragm.
1-Place slide on the stage and bring to focus.
2-Stop down the Field Diaphragm.
3-Focus the Condenser
4-Open the Field Diaphragm to the edge of the field and center using the condenser centering controls.
5-Adjust the Condenser Aperture Diaphragm to about 2/3 of full open.
Why do you use a coverslip for 100x oil objective?
To protect the specimen
True or False? For bright field microscopy, the refractive index of crown glass, used in microscope lenses, is 1.515; therefore, mineral oil, with a refractive index of 1.460 would make suitable immersion oil.
False
After the light source is centered, what is the next step in Kohler Illumination?
Close the field diaphragm
How does a CCD create pixels?
Using a grid of photo cells that sense light intensity
Various types of objectives can be used in the photographing of banded chromosomes. However, for optimal results objectives must correct for multiple wavelengths of light, curvature of field and have a high numerical aperture. If you were going to photograph a metaphase plate which of the following objectives would give the best result?
PlanApo 100/1.4 Oil 160/0.17
When moving from 10x to 100x you notice the image is blurry. This is because?
There are 2 cover slips, or the slide is upside down.
A tech looked at 10 different slides and printed different images for each. An artifact was found on each printed picture. What should be done?
Clean the printer.
What affects resolution?
NA of objective
NA of condenser
Refractive Index of the mounting medium
What is the thickness of the coverslip that can be used with the following labeled objective?
170/.17/oel/100/1.32
.17
What is a phase microscope used for?
Accessing amniotic fluid confluency
What is the function of the field diaphragm?
Controlling area of illumination
For photomicrography of chromosomes, I want the best resolution possible, therefore I will always use:
An objective with a N.A. of 1.3 to 1.4
Why cells cannot be observed under 100 x objective lens when slide is upside down?
In very short working distance, if the thickness of the cover glass exceeds the working distance then it is impossible to focus on the object.