Medtronic MYOTHERM XP Cardioplegia Flashcards
0
Q
Blood Line
A
1/4” or 0.6 cm
1
Q
Delivery Ratio
A
4:1
2
Q
Cardioplegia Line
A
1/8” or 0.3 cm
3
Q
Bellows
A
Stainless Steel
4
Q
Recommended BF Rate
A
up to 500 ml/min
5
Q
Filter Screen
A
150 microns
6
Q
HE Volume
A
44 mL
7
Q
HE Blood Contact SA
A
0.06 m2
8
Q
Inlet Connector
A
1/4” or 0.6 cm
9
Q
Outlet Connector
A
3/16” or 0.5 cm
10
Q
Pressure Monitor / Pressure Relief Ports
A
Standard Female Luer Lock
11
Q
Waterports
A
Hansen Series 3-ST Quick Disconnect
12
Q
Water phase pressure limit
A
40 psi
13
Q
Blood phase pressure limit (post prime)
A
750 mm Hg
14
Q
Max water temp
A
42 C