Measurement of Hearing Aid Performance Flashcards
If all seals are not sealed when testing in a couple box what will happen?
excessive gain at ~500Hz
Problems with measuring uncomplicated signal like pure tone or broad band noise
noise reduction kick in to reduce gain
feedback reduction kick in to reduce output
measures at different frequencies made sequentially may differ from those made simultaneously
program change may result in some aids with these types of signal (e.g. switch to music program)
What is the ICRA signal?
Artificial speech signals have been developed for the International Collegium of Rehabilitative Audiology
Benefit is is that it is somewhat standardised unlike proprietary modulated speech noises
What are the two main measurements made in a test box
The Gain Frequency Response
-This is the output measured with a 60dBSPL input signal (Or 50dB for a power aid). The gain at any frequency can then be calculated by subtracting the input SPL from the output SPL
The OSPL90
- Both the IEC and ANSI standards specify that hearing aid maximum output should be measured using a 90 dB SPL input signal, and both standards now use the term OSPL90 to describe the measurement. - When the hearing aid output has reached its limit for any input signal, it is said to be saturated. - Hearing aids with a steeply rising response will often not be saturated at low frequencies. In such cases, the measurement will underestimate the true maximum output of the hearing aid at low frequencies. - If you want to see saturation across the whole frequency range then you need to have everything on max gain, not PPL - The vertical axis of the OSPL90-frequency response graph is always shown in dB SPL - Do not take this in clients ear
Two commonly chosen sets of levels are
50,60,70,80and90dBSPL or 50,65, and 80
What is the difference between gain-frequency response and input-output functions
Gain-frequency response shows the gain (or output level) at all frequencies for one input level, Input-output function shows the output level at all input levels, for one frequency or for one broad-band test signal
What is the difference between responses (e.g. REUR) vs. gains (e.g. REIG)?
For any of the “Responses” (REUR, REAR, etc.) the figures you record are simple measures of the SPL in the ear canal.
For any of the “Gain’s” (REAG, REIG, etc.) the figures you are shown will be “total SPL – input level”.
Where should the probe tip be placed and why?
Within 6mm of the eardrum to eliminate error caused by standing waves
Information gained from IG measurements
- Amplification characteristics of the aid.
- Acoustic modifications to the aid.
- Acoustic effects of coupling the aid to the ear.
- Loss of normal ear canal resonance.
- Head/pinna effects
It is the REAG provided by the hearing aid, minus the REUG provided by the unobstructed ear
If the person’s ear canal and concha provide 12dB of gain at 2kHz, and the REAG (the SPL at the eardrum relative to the SPL outside the head) at this frequency is 30dB, what is the REIG?
18dB
Positioning of client for REIG
0.5-0.75m from speaker
At angle of 0-45degrees
What does a typical REUG response look like?
Fairly flat for the low frequencies and then a peak somewhere between 2.2kHz and 3.2kHz between 12 and 22dB above the low frequency plateau
Functional Gain
Aided thresholds = Unaided thresholds
Benefits of Functional Gain
Measures the clients aided hearing ability
Can be used in infants and children who might not sit still enough for REIG
Useful for profound losses - whether they are likely to get much benefit from acoustic cues in speech
Potentially better effectiveness because it is behavioural
Disadvantages of Functional Gain
Ambient noise floor can effect results
Needs cooperation from the client
Measured in 5dB steps so not heaps accurate
Only measures a small number of selected frequencies
Tells us how they respond to soft sounds not speech sounds
May cause feedback cancellation