Mass Spec Flashcards
what is mass spectrometer
instrument used to define the covalent structures of substances (biomolecules & proteins) by ionizing, separating, and detecting molecular and fragment ions according to their mass-charge ratios (m/z)
what make mass spec a powerful analytical tool
- very sensitive: able to produce structurally informative data from tiny amounts of starting material (eg. femtomoles or less)
- Able to obtain structurally informative data from complex mixtures of samples – don’t need highly purified, homogenous samples (eg. blood plasma samples, urine extracts, perfumes, protein digests etc)
what are the 3 key components of a mass spectrometer
ion source: convert biological samples into gas phase ions
mass analyser: Separate gas phase ions based on m/z ratio
detector: detect and quantitate ions (see which species are high or low in abundance)
what is the most abundant peak interpreted as
- Most abundant peak is given a value of 100% intensity
- other components are expressed as a relative percentage of the 100% peak
describe hard ionization technique + eg
- Generates enough energy to convert the biological samples into a gas phase ion but will usually have an excess of E from the ionization process, leading to fragmentation of the biological sample
eg. E Impact (EI) - can only ionize small molecules, 1-1000 Da
describe soft ionization technique + eg
- Enough E is provided to ionize the biological sample but no huge excess E which leads to fragmentation of the sample
- can ionize much bigger biological molecules ( larger protein complexes, not just individual proteins, and larger DNA fragments, polysaccharides, and biological polymers)
eg. Electrospray Ionization (ES or ESI)
– peptides, oligosaccharides, proteins, greater than 500,000 Da
eg. Matrix Assisted Laser Desorption Ionization (MALDI) – peptides, proteins, DNA, up to 500,000 Da
describe e impact ionization (EI)
- Performs ionization from gas phase - samples must be in the gas phase before EI
- Sample is introduced into the source by heating it from a probe tip until it evaporates or from an on-line gas chromatography
- Difficult to convert large biological molecules into the gas phase without causing decomposition (esp. if large hydrophilic proteins). Therefore, can only ionize relatively small molecules (esp. hydrophobic molecules = easy to convert)
explain the process of EI
- Gas phase sample is bombarded by a beam of high E electrons coming from heated rhenium or tungsten filament (energy = 70 eV)
* Can use magnets to focus the e- beam and maximize chances of inducing ionization - The e- beam then comes close enough to an outer valence e- of the sample to repel it (due to same – charge), excising the e- from the outer valency orbital of the molecule
* So Ionization occurs by loss of an e- to give M+. (a radical cation)
* Can only produce a single + charge on the sample - 70 eV in the e- beam will produce a great excess of E since an average bond E is only 5 eV, leading to bond breaking and fragmentation of the biological molecule (Most of the molecular ions will decompose into fragments via uni-molecular reactions)
- Now that the ions have got a charge, can start to control the movements of gas phase ions in the mass spec by introducing plates with a + charge for repulsion or - charge for attraction
* eg. + charge on ion repeller (back plate) to repel the M+ into the mass analyzer
- Not an efficient process – only a minority of molecules in sample will actually undergo EI
describe MALDI
-E comes from firing pulses of laser light at the biological sample
* Ionization from solid phase
explain the MALDI process
- Sample is embedded in a low molecular weight UV-absorbing “crystalline” matrix. The matrix has absorption maximum near the wavelength of the pulsed laser used to ionize the sample
* Sample is dissolved in a solvent and mixed with the matrix. The mixture is placed on a metal target where they will dry and co-crystallize, producing a 3D solid crystal lattice made up of the sample and the matrix - Introduce the metal target into a mass spec under vacuum & fire pulses of laser at target
- The matrix will absorb the laser pulse and enough E will be transferred to the sample to ionize it (& generate gas phase ions)
- In every MALDI experiment, will generate both + and – charged ions
- Once have charged gas phase ions, can now control their movement
* But can only analyze 1 charge state at a time
* SO depending on which charge ion we’re looking at, can put a charged back plate to repel the ions into the mass spec
describe MALDI lasers
-Most MALDI uses UV lasers: λ = 320-360 nm
Older version = nitrogen gas UV laser
More recent = solid-state UV laser
- Adv: able to fire more pulses of laser E at a quicker repetition = generate more ions = able analyze more samples in a fixed time frame
describe MALDI matrices
- Low molecular weight organic molecules
- Have double and triple bonds and conjugated ring system – act as chromophores for absorbing E from laser and allow E transfer process for ionization
alpha-cyano-4-hydroxycinnamin acid (aCHCA) - peptides and proteins
2,5-dihydroxybenzoic acid (2,5-DHB)- carbohydrates
what were early issues with MALDI
- Ions of the same m/z ratio coming from the target have diff. speeds when entering MS. This is due to uneven energy distribution by the laser pulse, depending on where in the crystal the ion is formed
- Some molecules will be nearer to the surface & will obtain higher E transfer from the laser while some = buried further in the crystal lattice, receiving much less laser E
what were some improvements for MALDI
Delayed Extraction
* Molecules nearer to the surface with higher E will move further from target plate after ionization
* Buried molecules with less E will move a shorter distance
- directly after ionization, ensure that no back potential is placed on the target to repel the ions into the analyzer
- Wait a few milliseconds
- SO that when the back potential is applied, slower ions (nearer to the potential) will receive more E and will be accelerated more to catch up with the faster ions (receive less repulsion from potential because they are further away)
* This compensates for the diff in ionization efficiency caused by ionization from a 3D crystal lattice
what is the advantage of electrospray ionisation (ES)
- Adv. = ionization from liquid phase (most bio. molecules = prefer to be in liquid phase)
explain the ES process
- Dissolve sample in a suitable solvent in a narrow glass needle that is coated with gold/ pallidum/ platinum at the tip. Depending on what aa is present, pH/pKa of solvent will give proteins in solution a charge
- Introduce the needle into the electrode
- A high voltage (3-4 kV) is applied to the tip (generate a strong electrical potential)
- A back pressure is generated so the sample will emerge from the tip as an aerosol of highly charged tiny droplets (containing peptides, proteins)
* Droplets = charged due to the pH, pKa of solvent, but also the high electrical potential - Drying gas (nitrogen) that flows around the outside of the needle is then applied to the droplets -this evaporates the solvent from the droplets
- Because solvent is evaporating, the droplets will get smaller and smaller, causing the charged ions to move to the surface
- At a critical point (Rayleigh limit) the same charged ions are forced too close together so the small droplets explode into ever smaller droplets
- Process continues until all the solvent is rid of, and left with gas phase ions
- Once have gas phase ions, can control movement into mass spec analyzer
* Can generate both + and – charged ions - only analyze 1 charge state at a time
* Manipulate by separating out only the charge we’re interested in (increase sensitivity)
What is the flow rate ES sources operate at
Early ES sources operated at flow rates of a few microliters/min (too much is released & need to keep adding back samples to generate the data)
Newer sources = nanospray sources - operate at flow rates of 10-30 nl/min
* Can have starting volume of 1 microliter and leave it spraying for several minutes
* Allow for many mass spec. experiments on a single sample loaded
* NanoES is much more sensitive than ES
* Sample quantities are typically in the subnanogram range
what does ES allow to do in addition
introduce the sample using high resolution Liquid Chromatography separation step prior to MS - very powerful method for analyzing complex mixtures
* Can inject sample directly from liquid chromatography into MA to generate mass spec data (because ionize from liquid phase) SO ES = favored for analyzing proteins and peptides
what are mass analysers and what do they do
separate biological molecules by m/z ratio:
* Quadrupole
* Time-of-flight (TOF)
* Ion trap
* Orbitrap
Depend on charge of gas phase ion for controlling their movement in electromagnetic field
what are 3 key performance characteristics of mass analysers
- Upper mass limit: the biggest molecule that can be successfully separated based on its m/z ratio
- Ion transmission: out of all ions produced in the source, how many will pass through the mass analyzer and get detected
- Resolution: how good is the mass analyzer at separating ions with very similar m/z ratio
describe a quadrupole mass filter
- made of four parallel rods
- Each diagonal pair of rods is connected electrically
- The electrical field is obtained by the application of a voltage made up of a Direct Current component (to one pair) and a Radio Frequency potential (to the other pair)
-Can change the voltages on the electrode to change the strength of the quadrupole field
describe the process of quadrupole mass filters
- Gas phase ions are directed out of the source into the quadrupole into the space between the 4 electrodes
- Depending on the m/z ratio, some ions will be in harmony with the quadrupole field and will travel straight through the quadrupole to the detector & a signal is obtained.
- Some ions with a diff m/z ratio will not be in harmony with the quadrupole, so they will get attracted to one of the electrodes and be annihilated, thus will not reach detector and will not get a signal in the mass spec
* If use a calibrant with a known m/z ratio, can produce a calibration curve which links m/z ratio to a specific quadrupole field
* SO for a quadrupole experiment, can choose a m/z range (ex. 200 – 2000) & quadrupole will allow ions to pass through to reach the detector, one m/z at a time as the entire mass range is scanned
what is the quadrupole performance criteria (lowest of all 4)
upper mass limit: able to analyze m/z ratio up to only 4,000 Da
* low sensitivity: due to scanning over a certain mass range & annihilation of ions, cannot detect every ion produced in the source (lose ions all the time)
* Relatively low resolution (unit resolution to about 3,000)
what are the pros of quadrupole
- Low cost
- Rapid scanning (robust)
- Small
- Ideal for Gas Chromatography-MS (EI) & widely used for electrospray-MS
describe ion trap analysers
operate on a similar principle to quadrupole analysers but do not operate as a filter
* also generate a quadrupole electromagnetic field but instead of having 4 electrodes on the same plane, they are in a sandwich arrangement
-composed of a ring electrode in the middle with cap electrodes (enter and exit electrode) on each end
describe the process in ion trap analysers
- Ions from the source will pass through the entrance electrode
- Ions are then trapped within the electrodes by the electromagnetic quadrupole field
- The trapped ions will start to circulate (spin around inside the ion trap)
- Next = alter the strength of the electromagnetic field
* Depending on the m/z ratio, some ions will receive an E boost, their trajectories will increase, and they will exit through the exit cap, appearing as a signal on the detector
-Mass spectrum is produced by scanning the RF voltages to eject all ions through the end cap
-Must calibrate with a standard known m/z ratio (know which voltage is used to eject which m/z)
what is the performance criteria of ion trap
- Has higher upper mass limit, better ion transmission, better resolution than quadrupole
what is the limitation of ion trap
relative small space
-if you produce a lot of ions, will not be able to capture and track them properly.
-some will collide into each other or w/the sides of the electrode, losing them in the analysis
how to overcome the ion trap limitation
high performance on trap = linear ion trap
-ions are trapped over a larger linear volume which helps overcome problems of ion interference and increases storage capacity
-get a bigger signal in MS: ions have more room to rotate around each other, don’t collide w/each other or sides = better sensitivity = better ion transmission
what is an orbitrap
-ions trapped by static electrostatic field
-electrode made up of an outer barrel electrode and an inner spindle electrode
-static electrostatic field is generated between the spindle and barrel electrodes
how does an orbitrap work
- ions are introduced into the orbital electrode
- because of the EM field, ions start to rotate around the spindle electrode, not just statically, but also horizontally (axial direction)
- ions w/different m/z ratio oscillate along the axial direction of the spindle electrode at a different rate
- FT converts time-domain signal to m/z
- must generate a calibration curve with a calibrant with a known m/z ration which relates a certain m/z ratio to a certain frequency of ion oscillation along the axial direction
what is the performance criteria of an orbitrap
-best performing of all 4
-highest mass accuracy (analyze up to v.high Mw molecules)
-highest resolution
-highest ion transmission (almost 100%)
What is a TOF analyser
an evacuated metal tube
explain the TOF analyser process
- ions from source will hit the evacuated tube where there is no electrical potential and travel the distance to the detector at another end
- Ions are separated by differences in velocities as they move in a straight path towards the detector
* Ions with a small m/z ratio = higher velocity
* bigger ions with larger m/z ratio = lower velocity
* Calibrate with calibrant of known m/z which relates m/z ratio to the velocity (amount of time it takes for an ion with a certain m/z to travel to the detector)
performance characteristics of TOF
measuring ions travelling down a tube so:
* “unlimited” upper mass limit mass range
* Close to 100% ion transmission – help sensitivity BUT: early analysers had poor resolution and mass accuracy
TOF limitations
Assumed that ions with the same m/z ratio will have the exact same E when they enter the flight tube (similar issue to MALDI)
* Ions with same m/z but diff starting E will travel at diff speed and hit detector at diff time, affecting mass spec quality
how to solve TOF limitation
a reflectron = corrects for the effects of unequal kinetic energy distribution of ions and gives a longer flight path
- an ion mirror that reverses the direction of travel of the ions
- ions of greater KE (higher V) penetrate further into the TOF before being reflected so have a longer flight path
- ions with lower starting E, initially slower = catch up by travelling a shorter flight path
-able to do so because they have a charge we can manipulate; put ion mirror of the same charge at the end of the flight tube to repel it for a 180º turn - compensates for the E difference so now ions w/ same m/z ratio hit the detector at the same time, giving a better spectra quality
-linear analysers have low resolution (<1000); reflectron gives much higher resolution (>3000)
what do ion detectors do
all detectors will amplify the signal to give higher sensitivity of detection
explain PM-photomultiplier
detects photons
1. ions strike a dynode, results in e emission
2. e then strike phosphorus screen, releases burst of photons
3. photons pass into multiplier where amplification occurs in a cascade fashion
4. number of photons is converted into an electrical signal, showing as a peak in the MS chart
explain EM-e multiplier
detects electrons
1. ions striking the dynode, will emit secondary electrons
2. amplification occurs when they continuously strike the next dynodes resulting in a cascade effect that produces more and more secondary electrons
3. e- are then converted into an electric signal and converted into a peak in the spectra
explain MCP (micro-channel plate array detectors)
- for simultaneous detection of multiple m/z values
- have hundreds of electron multipliers arrayed over a surface
explain the combinations of all components
- EI sources most common w/quadrupole analysers = cheap and small but low resolution
- Quadrupole = low performance but that is all that is required due to small generation of m/z ions in EI source
-GC-MS is widely used (a gas chromatograph (GC) coupled to MS); good for analysing low Mw non-polar compounds, metabolites - MALDI sources + TOF analysers
-MALDI-TOF = “mass fingerprinting” complex mixtures of polymers - ES sources most common w/ triple quadrupole (although ES can generate millions Da ions), ion trap, orbitrap or Q-TOF instruments (hybrid instrument)
-ES is especially useful for multiply charged molecules ex. peptides and proteins
explain hybrid experiments
- have 2 or more analysers in tandem ex. the Q-TOF
- Other common arrangements are TOF-TOF (usually combined with a MALDI source) and linear ion trap-orbitrap.
- Hybrid instruments are used for MS/MS experiments (2D MS analysis)
give examples of hybrid instruments
4800 MALDI TOF/TOF Analyzer - Femtomolar sensitivity
MALDI AXIMA Resonance QIT
* Quadrupole, Ion Trap, TOF Mass Analyzer system
* ~1000 resolving power
Orbitrap Ascend TribridTM Mass Spectrometer
* Quadrupole, Multipole, C trap, orbitrap, linear ion trap
* Sub fmol sensitivity
* High resolution capability, 1,000,000
what type of data does MS contain
- Molecular ion signals
- Fragment ion signals
what is a molecular ion
- Made from intact biological molecule
- major type of ions that is generated from soft Ionization techniques - MALDI, ES
explain real molecular ions
- generated by Electron Impact (EI)
- Because molecule lost a single electron, directly generating singly + charged radical ions (M+.)
-Their m/z will be equal to mass of molecule (m/1- because mass of e- lost is small and charge = 1)
explain pseudo molecular ions
Generated by MALDI/ES
-Because charge is generated by adding a counter ion during the ionization process
-if observing + charged species, need to add + charged ions (ex. protons, metal
ions ex. Na+, K+, Li+ or if observing – charged species, can remove a proton)
explain ions generated by MALDI
-Can get singly charged cations and anions
-m/z = mass of peptide +/- mass of added/ removed ion
eg. if peptide = 100
[M+H]+ = (100 + 1)/1 = 101
[M+Na]+ = (100 + 23)/1 = 123
[M-H]- = (100 - 1)/1 = 99
explain ions generated by ES
-Can get multiply charged cations and anions i.e. [M+nH]n+, [M+nNa]n+, [M-nH]n-
- m/z = (mass of peptide +/- mass of added/removed ions) / charge
eg. if peptide = 100
[M+2H]2+ = (100 + 2)/2 = 51