Mammography Flashcards

1
Q

Mammography units that can perform digital breast tomosynthesis are most likely to have a tube target composed of ________.

A

Tungsten (W)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

For tomosynthesis units, ____________ is (are) commonly used as the x-ray tube filter material(s).

A

silver, aluminum, and rhodium

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

The latest American College of Radiology (ACR) mammography accreditation phantom includes what objects?

A

6 fibers, 6 groups of specks, and 6 masses

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How does the breast radiation dose from breast tomosynthesis compare to that of conventional mammography?

A

About the same

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

The average glandular dose when scanning the ACR digital mammography phantom must be less than _____.

A

3 mGy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

The maximum compression force for the inital power drive in mammography must be between ____ and _____.

A

11 daN, 20 daN

OR

25 lbs, 45 lbs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

According the MQSA, the tolerance for the x-ray / image receptor alignment is

A

2% of the SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are the minimum phantom scoring requirements for the old ACR mammography phantom?

A

4 fibers, 3 spec groups, 3 masses

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What are the minimum phantom scoring requirements for the old ACR mammography phantom?

A

2 fibers, >=3 speck groups, 2 masses

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

The ACR digital mammography phantom simulates what?

A

A “standard breast” of 4.2cm thickness and 50% glandularity.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly