LVADS Flashcards
after load sensitivity of cf LvAD
CF LVAD flow highly sensitive to after load. therefore need a narrow range of pressures
HEAD CURVE
of axial vs rotary LVAD
Pump head pressure (dP across the pump) vs flow.
Centrifugal pumps have a flat head curve — operate over a wide range of flow for a small dP.
- results in some pulsatility
Effect of flat head curve on low flow conditions
steeper head curve of axial pumps responds to an increase in SVR by increasing the pressure generated across the pump, limiting the decrease in flow by increasing outlet pressure.
At low-flow conditions, this retains the capability to enforce high blood pressures with adequate LV volume.
** However, it also results in high inlet suction that in a low LV volume state can potentiate arrhythmias, create suction events, or lead to hemolysis.
formula for PI for HM 2
he formula for calculating a PI for the HeartMate II (PIHM II) [(flowmax – flowmin)/flowavg] 10.