Linac Faults Flashcards

1
Q

What must you consider regarding patient safety?

A
  • Is patient comfortable
  • Is it possible with patient on the couch
  • Is patient wearing uncomfortable immobilisation equipment
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What if it is a gun fault?

A
  • No gun, no electrons, linac will absolutely not work
  • Where is the gun, what else will be affected, what other components?
  • Not easy to get a new gun, maybe ONE spare
  • Replacing the gun is a major task as it disrupts multiple parts of the linac and all must be checked
  • Machine may be out of use for up to two days
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What if it is a vacuum fault?

A
  • ion pumps must be switched off before looking for vacuum fault.
  • Must then check each seal and find the leak.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What if there is a steering fault?

A
  • If the electron beam hits the target offset, it will produce a beam profile but they will be incorrect
  • A monitoring system will stop the machine (ion chamber)
  • Physics and engineers will need to be consulted
How well did you know this?
1
Not at all
2
3
4
5
Perfectly