Lecture 2 Flashcards
What are the three things we need to record far-field responses?
- Spatial summation
- Synchrony
- Open-Field orientation
What imaging technique uses the right hand rule?
MEG
How much is being measured through ABR?
± 0.2 µV
What range can computer’s pick things up in? What happens if things are outside this range?
- A range of ± 10 V
- Anything greater than this will be clipped (heavily distorted)
How much amplification would be appropriate?
100,000 x .2 µV = 20 mV
What does ABR come packaged in? How much noise are we often dealing with?
Noise (often dealing with 20µV of noise)
How much is being measured in MLR?
± 0.4 µV
How much is being measured in LLR?
± 4.0 µV
MLR is spread over ____ and LLR is spread over ____
50ms, 500ms
How to prevent shocking the system
- Dryness is most risky—can use humidifier if air is dry!
- Leather-souled shoes are better than rubber-souled shoes (insulating).
- When in doubt, always discharge before touching equipment by touching metal (e.g., table leg).
Explain common-mode voltage (line noise)
- Wall voltage = 120,000,000 µV (300,000,000 times larger than ABR)
- If near any source of electricity, ABR can pick this up (need to position correctly)
- Almost always have a bit of common-mode voltage
What are strategies to minimize common-mode/line noise before doing test?
- differential amplification
- balanced impedances
- electrode care (placement, braiding)
What are strategies to minimize common-mode/line on the computer?
- averaging
- filtering
Explain differential amplification
Differential amplification is amplifying the difference between two things
What are the the two electrodes we are finding the difference between?
- Active (non-inverting)
- Reference (inverting)
response = (active) - (reference)
What is the ground?
the person
What does the ground do?
- ground electrode provides the amplifier with a ‘reference voltage’ to prevent amplifier drift (to stay in the voltage range of amplifier; active & reference)
- also helps with common-mode rejection
How to reject common mode?
- You need a balanced amplifier
- common mode rejection ratio (CMRR)
What is the common-mode rejection ratio?
10000/1(80dB) to 100000/1(100dB)
What is coupling?
Connecting the electrodes to the peron
What do you need for coupling to take place?
- a conducting medium (skin doesn’t contact the electrode directly)
- a paste or jelly is used
- electrolyte should come into contact with subcutaneous tissue (exfoliation)
What is the impedance of the epidermis (outer layer) before being exfoiliated?
25 kΩ (x2 = 50)
amplifier input impedance must be much higher than sum of all skin contact impedance
skin potentials more likely with ____ impedance
high
What is the target impedance?
below 5 kΩ
What is the electrode made out of? What filter does it use?
- Metal immersed in electrolyte, ion exchange, two layers (opposite charge) building-up adjacent to the metal
- this acts as a high-pass filter (removes low frequencies)
- depends on particular metal
What types of electrodes are there?
- Silver/Silver-Chloride (most common for high end systems)
- Silver
- Tin
- Gold (often in the clinic)
- Platinum
High-impedance amplifier reduces ____
Distortion
For good CMR, it is also necessary to have balanced ____
amplifier-person connections
electrode impedances should be similar for all electrodes - within ____
1 kΩ
Explain eletrode-skin impedance
1. what two things are essential
3. why does it need to be low
- Must be balanced (within 1 kΩ) for good CMR
- Must be much lower than input impedance of amplifier (usually below 5 kΩ)
- Should be low to reduce effects of skin potentials
- can only be completely removed by puncturing skin
- sweat is an enemy, keep subjects cool
- generally low frequency - can be removed with filtering
Explain the two main things with electrode placement
- wires should be away from devices with current (e.g., ER-3 earphones)
- do not dangle in face!!
- requires some planning
- generally electrodes go towards back of the head - electrode wires should be braided
- this minimizes current induction on the wires
- very inexpensive way to improve recording
What is electrode montage?
where we place the electrodes