Lecture 1-3 Flashcards

0
Q

Selenium^(c,k) means?
What is INAA?
What is GFAAs?

A

Instrumental analytical methods used to analyze SRM.
Instrumental neutron activation analysis.
graphite furnace atomic absorption spectrometry - make sure two methods get the same [].

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
1
Q

Analyte [] in a sample: Major ____; minor ____; trace ____.
SRM stands for ? NIST stands for ?
1ppm = ____
1ppb = ____

A

> 1% 0.1-1% <0.1%
Standard Reference Material
National Institute of Standards and Technology
1mg/L 1ug/L

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

s = ____; s^2 = ____; relative s (%) = ____.
Measurements with ____ s more ____ than those with a ____ s. Does not imply _____ ____. Single most important characteristic ofany result from analytical measurements: statement of its ____.

A

s= standard deviation s^2 = variance s(%) = 100*s/x (bar)
small precise larger
greater accuracy
unvertainty

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

One method to determine __ analyte at the same time: ____.
Separation in mobile phase of two compounds, similar -> caffeine vs. theobromine, one has more ____ -> distinguished. It can detect by ____ ____ -> ____ ____ full; -> molecule passing ____ light. ____ determines quantification.

A

2 chromatography methyl group
light intensity
mobile phase dimer Intensity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Peak height of each peak ~____ of each compound. Need to standardize ____ response to known ____ of ____ analyte. For pure theobromine and caffeine with ____ ____ with 10,25,50 and 100ppm. On the graph, y-axis is ____ and x-axis is ____.

A
quantity
detector    quantity    each
standard solutions
y = peak height (cms)
x = analyte [] (parts per million)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Method of least squares: often use in context of a “____ ____” to fit best line through data points. To ____ magnitude of + and - deviations -> ____ all deviations.
Two assumptions: 1. error in __ value > than that of __ values. 2. ____ s for all data.

A

calibration curve
minimize square
y value x value
same

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Determine protein [], (protein usually colorless):
1. ____ solution preparation, ____ solution of ____. (covering range of __ expected from ____). 2. ____ ____ ____ of the ____ solution from each measured absorbance to get ____ ____. 3. make a ____ of ____ ____() vs. ____ of protein(). Good practice: at least _ replicates of sample - make a graph of data with _ . -> can reject bad data or repeat measurements.

A

Blank standard analyte [] unknown
Substrate average absorbance blank corrected absorbance
graph corrected absorbance (y) amount (x)
2 6 []

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

QA quantity assurance -> do to ___ the analytical results in terms of ____ and ____. International measurement evaluation program (Belgium) - __ ____ in water measurement.

A

verify accuracy precision

Pb []

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

UV-vis-spectrophotometer measures ____ ____ ~ ____ of protein. Absorbance from black:

  • ____ of reagents
  • ____ in reagents
  • ____ species
A

light absorbance amount
color
impurities
interfering

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Basics of QA: 1. ___ data, ____ measurements 2. ____ data: __’s derived from applying a ____ procedure to ____ data. 3. ____: ____ reported after ____ analysis of ____ data: __, _, ____ ____.

A

raw individual
treated []’s calibration raw
results quantities statistical treated
x(bar) s confidence interval

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Analytical use objectives: 1. state ____ for which results will be used (ex customer needs) 2. ____ of data want to collect and ____ to collect. 3. how ____ and ____ needs to be; speed over time, cost. Example: treatment of oil -> decrease ____ in soil to ____ ____ by ____ activity.

A

purposes
kind ways
precise accurate
[hydrocarbon] threshold value microbe

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Define “specifications”:

A
  • how good must numbers be and what precautions in analytical procedure.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Soil sample: ____ on surface of soil, is ____ than the deep centre of the pile -> several samples needed from different ____; -> full container of soil packed, no ___ within; -> label it. (be careful of ____/____); -> not at ____ temperature, ice chest. ____ chew down hydrocarbons. Check the ____ in the container, if not right, no need to analyte.

A

[hydrocarbon] less
position air volatilization/ degradation
room Microbes
temperature

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Test ____ of the device before analyzing. - use ____ material; - make sure ___ has the right __ as on the label, result acceptable close. - reagent ____ needed. (highest or introducing ____), tolerance of ____; model using.

A

precision
standardized SRM []
purity contamination apparatus

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Rate of false results: even well executed procedures sometimes produce some false conclusions: statistical nature of sampling and measurement.
False positive : [] ____ legal limit when, in fact it’s ____ limit;
False negative: [] ____ legal limit when, in fact it’s ____ limit.

A

exceeds below

below above

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Selectivity : able to ____ analyte from other species; detection limit must be ____ than expected []. Acceptable blank values: 1. ____ blank: sample containing ___ component except _____ taken through all steps of ____ ____ (____ from sample measurements.) 2. ____ blank: similar to ____ blank, but has been exposed to site of ____. This can tell us if analyte inadvertently picked up by exposure to field conditions.

A

distinguish lower
Method all analyte analyte procedure subtract
Field method sampling

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Recovery of fortification: ____ recovery: sometimes response to ____ can be ____ by something else in sample (____ effect, ____ else in sample). Spike = ____ quantity of analyte added to ____ test if response as expected from ____ ____.

A

spike analyte decreased matrix everything
known sample calibration curve

17
Q

Calibration checks: ____ number of samples -> check ____ continues to work properly.
Quality control samples (____ test samples/____), sample provided to analyst as ____ obtained results compared to ____ values.

A

large instrument

performance blind unknowns known

18
Q

SOP stands for? Defines exactly what analyst has to do (____); must be ____ followed to prevent ____ results due to “short-cuts”.
____: 1. collect data which show that ____ procedure operates ____ specified limits. 2. do final ____ meet defined ___ ____.

A

Standard operation procedures
initial rigorously false
Assessment analytical within results use objective(s)

19
Q

API stands for?
____ ____: process of ____ analytical method acceptable for purpose.
API = ___ [], method requirements for regulatory submission: 1.____ 2. ____ 3. ____ 4. _____5.____ 6. ____

A

Active Pharmaceutical Ingredient
Method Validation proving
pure
1.method specificity 2. linearity 3. accuracy 4.precision 5. range 6. limit of detection (DL/LOD) and quantification (LOQ)

20
Q

Method specificity: ____ of analytical method to ____ analyte from everything else that ____ be in sample electrophoresis: substances ____ by differing rates of ____ in an electrical field.

A

ability distinguish
might separated
migration

21
Q

Linearity: how well is response ____to quantity of analyte; how well does ____ ____ follow a ____ line.

A

proportional
calibration curve
straight

22
Q

Target [] of API in drug formulation known-> test ____ of calibration curve with _ standard solution’s spanning range form __ to __ times of expected [].

A

linearity 5

0.5 to 1.5

23
Q

Each standard prepared and analyzed 5 times: 3X5= __ standards + __ blanks. One measure of ____ is the square of ____ ____ R^2. Another criterion: _-intercept of calibration curve (after ____ subtracted form each standard) should be close to _.

A

15 3
linearity correlation coefficient
y blank
0

24
Q

Accuracy (nearness to truth): 1. analyze an appropriate ___; 2. compare results from _ different analytical methods; 3. analyze ___ sample with a ____ addition of analyte (____ same as unknown); 4. ____ ____ of analyte to unknown. Because SRM + second analytical method often not available, ____ most common.

A
SRM
2
blank       known       matrix
standard addition
spiking
25
Q

Precision: how well ____ agree with one another (small _); autosamplers in ____ and ____ have improved precision by factor ___.

A

replicates s
chromatography and GFAAS
3-10

26
Q

Instrument precision: ____ when same quantity of sample repeatedly introduced (__ times). Interlaboratory: precision measure of ____ by different ppl in different laboratories, as [] of analyte ____ becomes poorer coefficient of ____/____ s: CV(%)=100*s/x(bar)

A

reproducibility >=10
reproducibility
decreases
variation/relative

27
Q

Range: [] interval over which ____, ____ and ____ are all acceptable. Linear range: __ range over which measurable response. Define __ interval for which: 1. ____ ____ R^2 >= 0.995 2. ____ recovery 100+-2% 3. ____

A
linearity, accuracy and precision
[] []
correlation coefficient
spike
interlaboratory
28
Q

____ of ____ (DL/LOD) and ____ (LOQ): smallest analyte [] that is statistically “____ ____” from blank (____). DL -> _% change of concluding that blank has analyte -> false ____ and __% chance that analyte ____ because ____.

A

Limits of detection quantification
significantly different noise
1% positive
50% absent signal

29
Q

Steps to follow: 1. measure n ____ (no analyte) -> y(blank) + s 2. ____ detectable signal y(dl) (___ ___ ___ y(dl) = y(blank) + 3s ) 3. prepare sample whose [] ~___ times DL 4. measure n replicate samples (n >= _) y(sample) 5. ____ signal: y(sample)-y(blank)~ m*[sample] (m is the ___ of linear calibration curve).

A
blanks
minimum
1-5
7
slope
30
Q

Minimum detectable [] = ____ ____ (DL) = _ s/m. Signal 3 X noise-> detectable-> no accurate measurement. Lower limit of quantification (LOQ) = _ s/m

A

detection limit 3s/m

10s/m

31
Q

Reporting limit: [] below which regulations says that a given analyte is reported as “___ ____”. Does not mean analyte is ___ ____! ___ times higher than __.

A

not detected
not detected
5-10 DL

32
Q

U.S. packaged food must state how much ___ fats present (heart disease). Reporting limit : ___ fats 0.5g/serving -> [] _ 0.5g/serving reported as 0. -> ____ serving size so [] less than 0.5g/serving.

A

trans

trans

33
Q

Robustness: ____ of analytical method to be ____ by small ____ changes in operating parameters (injection volume, detector lenda, room T, etc.).

A

ability
unaffected
deliberate

34
Q

Cause of ___ effect: other anions in ground water when sample composition complex and affects reading add ____ quantities of analyte to ____ and analyze.

A

matrix known unknown

35
Q

Standard addition: standard ____ substance as analyte
Internal standard: ____
Internal standard: instrument response varies slightly run to run? ->____ ____ useless. Add known amount of ___ ____ to unknown from analyte.

A

same different
calibration curve
internal standard

36
Q

Analyte signal area for X compared with that from internal standard S -> determine __ of X area under each peak proportional to __ of each species and ____ response to X to S. In practice we ____ solution of analyte with ____ amount of standard ->analysis.

A

[] [] linear

spike known

37
Q

Detector response to each component similar? If both ____ (X) and ____ ____ (S) have [] of 10mM -> area under analyte peak might be ___ times greater for X for S: ____ ____ (F) is ___ times greater for X than for S.

A

analyte (X) internal standard (S)

  1. 30 response factor (F)
  2. 30
38
Q

(a__)/(__b)= F(c__)/(__d)
When F=1, means detector responds ____ to X and S.
When F=2, means detector responds ____ as much to X than S. {X} and [S] are [] of analyte and standard ____ they have been mixed.

A
a= area of analyte signal
b= [] of analyte
c= area of standard signal
d= [] of standard
equal         twice
after
39
Q

Basics of QA = what we do to get right answer for purpose, 1.write ____ ____ (how will you use results) 2. ____ (how good do numbers need to be) 3. ____ (were specifications achieved).
____ ____: analytical method acceptable for intended purpose.
____ ____: analyte in sample where matrix affects response.
____ ____: response varies from run to run.

A
  1. use objectives
  2. specifications
  3. assessment
    Method validation
    Standard addition
    Internal standard