Last exam questions Flashcards

1
Q

Describe a fluorimeter, what parts the instrument is made out of and their function

A

Light source: The light source provides the excitation energy necessary to induce fluorescence in the sample. Common light sources include high-intensity LEDs, xenon lamps, or lasers.

Monochromator: The monochromator is used to filter the light source, selecting the excitation wavelength. This allows the user to selectively excite the sample at a specific wavelength.

Sample holder: The sample holder is used to hold the sample in place during measurement. It may be a cuvette or a microplate, depending on the type of sample being analyzed.

Detector: The detector is used to measure the intensity of the emitted fluorescence. Typically, a photomultiplier tube or a CCD camera is used for detection.

Filters: Fluorimeters may use filters to remove any scattered light or stray excitation light, which can interfere with the fluorescence signal.

Data analysis software: The data analysis software is used to analyze the fluorescence data and to calculate the concentration of the fluorescent molecule in the sample.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

For a fluorimeter, what substances can be analyzed and what advantages there is to florescent spectroscopy compared to absorption spectroscopy

A

Sensitivity: Fluorescence spectroscopy is a more sensitive technique than absorption spectroscopy, as it is capable of detecting much lower concentrations of molecules. This is because fluorescence measurements are made on the basis of the emission of photons, rather than the absorption of photons, and fluorescence signals can be amplified through signal integration over time.

Selectivity: Fluorescence spectroscopy offers greater selectivity than absorption spectroscopy, as it is possible to selectively excite specific molecules or groups of molecules using a specific excitation wavelength. This selectivity can be used to detect and quantify specific molecules in complex mixtures, such as biological fluids.

Structural information: Fluorescence spectroscopy can provide valuable information on the structure and conformation of molecules. For example, changes in the fluorescence emission spectrum of a protein can provide insights into changes in its structure or conformation.

Non-destructive: Fluorescence spectroscopy is a non-destructive technique, which means that the sample is not altered or destroyed during analysis. This makes it possible to analyze the same sample repeatedly, or to recover the sample for further analysis using other techniques.

A wide variety of samples can be analyzed such as nucleotides and proteins, chemicals and drug detection in samples, pollutions, certain metals etc

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Based on the chemical structure of paracetamol, explain why varying pH during UV absorbance spectroscopy may have an effect of the calibration curve (it has a NH group, O, CH3, HO and a carboxygen ring

A

At acidic pH values, the phenol group may become protonated, resulting in the formation of a positively charged species. This can lead to a shift in the absorption peak to a lower wavelength, which can affect the calibration curve. Additionally, at high acidic pH values, the amide group may undergo hydrolysis, leading to the formation of a degradation product that may also have a different UV absorbance spectrum compared to paracetamol, leading to further distortions in the calibration curve.

At alkaline pH values, the amide group in paracetamol may become deprotonated, resulting in the formation of a negatively charged species. This can lead to a shift in the absorption peak to a higher wavelength, which can again affect the calibration curve.

Furthermore, variations in pH can also affect the solubility and stability of paracetamol, which can further complicate the UV absorbance measurements. For example, at very high or very low pH values, the solubility of paracetamol may be reduced, which can affect the amount of sample that can be analyzed and the accuracy of the measurements.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

If the absorption ϵ, is 11,329 at 243 nm and the cuvette is 1cm and the result is 0.75 what are the concentration of paracetamol?

A

The Beer-Lambert law states that the absorbance (A) of a sample is directly proportional to the concentration (c) of the absorbing species and the path length (l) through which the light passes through the sample, according to the formula:

A = εcl

where ε is the molar absorptivity or molar extinction coefficient of the absorbing species.

Given the following parameters:

ε = 11,329 M^-1cm^-1 at 243 nm
l = 1 cm
A = 0.75
We can rearrange the Beer-Lambert law to solve for the concentration (c) of the absorbing species:

c = A / (εl)

Substituting the values given above, we get:

c = 0.75 / (11,329 M^-1cm^-1 x 1 cm) = 6.62 x 10^-5 M

Therefore, the concentration of the absorbing species is 6.62 x 10^-5 M.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

How do you calculate retention factor? Do question 3 A.

A

k = (tr - t0) / t0

where tr is the retention time of the compound and t0 is the retention time of an unretained compound or the dead time of the column.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

How do you calculate plate number? Do question 3b

A

N = L / (5.54 * (tR / w0.5)^2)

where L is the length of the column, tR is the retention time of a peak, and w0.5 is the width of the peak at half-height.

To calculate the plate number, the retention time and peak width at half-height must be measured for a peak of interest. The length of the column is a known value and must be entered into the formula.

The factor 5.54 is a constant that depends on the flow rate of the carrier gas and the temperature of the column. It is commonly used for columns operated at flow rates of 1-100 mL/min and temperatures of 20-30°C.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

How do you calculate resolution? How can we see if we have a baseline peak based on the resolution of two peaks? Do question 3c and d

A

The resolution (R) is a measure of the separation between two adjacent peaks in a chromatogram and is calculated using the following formula:

R = 2 (t2 - t1) / (w1 + w2)

where t1 and t2 are the retention times of the two peaks, and w1 and w2 are their respective peak widths at half height.

To determine if there is a baseline peak based on the resolution of two peaks, one can calculate the resolution using the formula above. A resolution value of 1.5 or higher indicates a good separation between the two peaks, with little or no overlap. A resolution value of less than 1.0 indicates that the two peaks are not well separated and may be overlapping, indicating the presence of a baseline peak.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

How can you tell based on 2 different chromatogram looking at the peaks which chromatogram had the thickest film the one with no baseline separation and lower peak heights or the one with baseline separation and higher peak heights? Why?

A

The one with high narrow peaks and seperation has thinner film since the analyte will interact less with the stationary phase and pass faster through the column which geaves more narrow and sharp peaks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What three column parameters can be changed to increase plate number N?

A

Column Length: By increasing the length of the column, the number of theoretical plates increases, leading to an increase in the plate number N. However, increasing the column length also increases the retention time and the analysis time.

Particle Size: Smaller particle size packing materials in the column lead to a higher number of theoretical plates and thus an increase in the plate number N. However, decreasing the particle size can also increase the pressure drop in the column, requiring higher operating pressure.

Column Diameter: By decreasing the column diameter, the band broadening and dispersion effects are reduced, leading to an increase in the plate number N. However, decreasing the column diameter can also increase the backpressure and reduce the sample loading capacity.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Draw the chemical structure of a common phase in gas crhomatographt and give trhe chemical namwe of this chromatography. Both the polar and non polar one

A

SiO2 and Si-O-Si-C18H37

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

How do you achieve gradient solution in gas chromatography? Give an example of such gradient

A

One example of a gradient solution in GC is a temperature program, where the temperature of the column is increased over time to elute different compounds from the sample. For example, if analyzing a mixture of compounds with different boiling points, a temperature gradient can be used to selectively elute each compound as the temperature of the column increases.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Give an example of a suitable stationary phase and mobile phase in isocratic elution with reversed HPLC

A

HPLC is C18-bonded silica as stationary phase and water and organic solvent such as methanol for example usually 50/50

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Give an example of the mobile phase composition can be changed in gradient elution reversed phase HPLC. How does the eluent strength change during the gradient? Explain

A

One can change the concentration of water and methanol to detemrine how fast the analyte shoudl move throught he column. The more it interacts with the mobile phase the faster it will go so increaisng or decreasing methanol to slow down or speed up.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

In reversed phase HPLC what will be the elution order for oxazepam and diazepam? Explain based on the chemical structure of compounds and their interactions with the stationary phase

A

Diazepam is more hydrophobic then oxazepam and will therefore elute after oxazepam. This is because oxazepam has an NH group and an OH group which diazepam doesn’t which makes the group less hydropohibic

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

In what three ways can separation factor be affected for HPLC

A

Changing the mobile phase composition: The separation factor is affected by the composition of the mobile phase, which can be adjusted by changing the ratio of solvents used in the mobile phase or the pH of the buffer solution. By changing the mobile phase composition, the separation factor can be adjusted to improve the selectivity of the separation.

Changing the column temperature: The separation factor is also affected by the column temperature. Increasing the column temperature can decrease the retention time and change the separation factor for some compounds, which can be useful for improving selectivity and/or speeding up the separation.

Changing the stationary phase: The separation factor is also influenced by the chemical and physical properties of the stationary phase. Different types of stationary phases, such as C18, C8, or phenyl columns, can have different selectivity for different types of solutes. Changing the stationary phase can therefore be a useful approach for optimizing the separation factor and improving the selectivity of the chromatographic separation.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

True or false:
HPLC: A column packed with large particles will have low plate number N

A

True.

The number of theoretical plates correlates with the particle size. The smaller the particle, the higher the number of theoretical plates can be.

17
Q

True or false:
HPLC: A decreased column internal diameter will have increase in plate number N

A

True

18
Q

In capillary electrophoresis sample induction is made by loop injection

A

False.

In capillary electrophoresis (CE), sample introduction is typically achieved by electrokinetic injection, where a voltage is applied across the capillary to induce the migration of sample components into the separation buffer.

19
Q

Helium gas is a common carrier gas in gas chromatography

A

True.

The most common carrier gas is helium since it provides good separations, is inert, and is safe to use

20
Q

Small molecules elute before large molecules in molecular exclusion chromatography

A

False. After sample has entered the column, molecules larger than the pores are unable to diffuse into the beads, so they elute first.

21
Q

GC: To achieve focusing by solvent effect the injector temperature must be 20-40 degrees below the boiling point for the analyte

A

True

22
Q

Write about the various abbreviations and describe them:
[MH+]
TIC
EI
MRM
HRMS
m/z

A

TIC: This abbreviation stands for the total ion current, which is a measure of the total ion signal in a mass spectrometry experiment. The TIC represents the sum of all the ions that are detected over the entire mass range, and it is used to evaluate the overall quality of the mass spectrometry data.

EI: This abbreviation stands for electron ionization, which is a common ionization technique used in mass spectrometry. In EI, a high-energy electron beam is used to ionize molecules in the gas phase, resulting in the formation of fragment ions that are characteristic of the molecular structure of the analyte.

MRM: This abbreviation stands for multiple reaction monitoring, which is a targeted mass spectrometry technique used for quantitative analysis of specific compounds. In MRM, the mass spectrometer is set to monitor the transition of a specific precursor ion to a specific product ion, allowing for highly sensitive and specific detection of the analyte of interest.

HRMS: This abbreviation stands for high-resolution mass spectrometry, which is a type of mass spectrometry that offers high mass resolution and accuracy. HRMS instruments are capable of resolving complex mixtures and providing detailed structural information about the analytes, making them particularly useful for the identification of unknown compounds.

m/z: This abbreviation stands for mass-to-charge ratio, which is a fundamental parameter used in mass spectrometry. The m/z value of an ion is determined by dividing its mass by its charge, and it is used to identify and quantify ions based on their mass and charge characteristics.

23
Q

We have a calibration curve with a straight trend line, y axis represent analyte area and x axis represent added concentration (ng/mL). Y=545.08x+6461.8 and R squared is=0.9949. The line starts at 0. What is the problem with this calibration curve?

A

The trend line shows that we start at 0 but the equation states that the starting value is 6461.8 which is not represented in the curve and doesn’t make any sense for what we are measuring nor the expected values. Therefore it’s something wrong with the equation and we need to fix it. 545.08 x is also not represented in the graph, and y axis shouldn’t be analyte are abut a volume

24
Q

We are developing a LC-MS method for determination of sertraline (305.0738D) in plasma. The method is gonna be used to determine bioavability of the substances after oral administration. Specify the criteria to ensure positive identification of sertraline and explain how to ensure correct identification

A

Accurate mass retention time, can determine elemental composition and determine the molecule. Look at retention time, isotope pattern and fragmentationj time since some compounds have the same elemental composition

25
Q

We are developing a LC-MS method for determination of sertraline (305.0738D) in plasma. The method is gonna be used to determine bioavability of the substances after oral administration. Suggest and motivate a mode (positive of negative) together with an ionization method for sertraline that are suitable for LC

A

Acid-Negative ionization-add an electron
Base-positive ionization-remove an electron

For the determination of sertraline in plasma using LC-MS, a positive ionization mode would be suitable. This is because sertraline is a basic molecule(since it has a NH that can give a proton) that has a pKa of around 9, which means it is more likely to ionize positively in the presence of a protonated solvent.

In terms of the ionization method, electrospray ionization (ESI) would be a good choice for sertraline. ESI is a soft ionization technique that produces a high abundance of molecular ions and minimal fragmentation. It is also compatible with LC, making it a popular choice for LC-MS applications.

With ESI, a nebulizer gas is used to generate charged droplets of the analyte solution, which then enter a high voltage field that induces the formation of ions. The ions are then directed into the mass analyzer for separation and detection.

26
Q

A triple quadrupole instrument can be operated in several modes, inspired by the figure below describe three modes and give examples of when they are used

A

Product ion scan (also known as MS/MS or tandem mass spectrometry): In product ion scan mode, the first quadrupole selects a precursor ion from the sample, which is then fragmented in the second quadrupole. The resulting fragments are then analyzed in the third quadrupole, generating a spectrum of product ions. This mode is used to identify and characterize the structure of a specific molecule.

Single reaction monitoring (SRM): In SRM mode, also known as multiple reaction monitoring (MRM), specific precursor-product ion transitions are selected to selectively detect and quantify a specific molecule or class of molecules. In this mode, the first quadrupole selects a precursor ion and the third quadrupole selects a specific product ion for detection. This mode is commonly used in targeted quantitative analysis of known compounds in complex samples.

Precursor ion scan: In precursor ion scan mode, the first quadrupole selects a specific precursor ion or class of precursor ions, and the third quadrupole scans for all resulting product ions. This mode is used to identify the presence of specific classes of compounds or to screen for unknown compounds in a complex mixture.

27
Q

Look at question 2b, what is paracetamols Lambda max and why is it important to know?

A

. λmax is an important parameter to know when analyzing a calibration curve because it’s the wavelenght we use to detemrine concentration, it can help to analyze unknown compounds and