KV Radiation Equipment Flashcards
1
Q
KV PDD (2)
A
-d-max at skin surface
- not penetrating
2
Q
KV Treatments (2)
A
- Good for skin treatments
- High superficial dose
3
Q
KV Scatter
A
low and medium energy x-rays readily scatter in air
4
Q
Applicators (3)
A
- used to achieve sharp penumbras and delineate treatment fields
- solid walls, not thick due to x-rays hitting at an oblique angle
-applicators extend to patient surface
5
Q
Lead cut outs (4)
A
- minimizes normal tissue irradiation
- customized for patient lesion
- lead thickness dependent on x-ray energy
- must be slightly larger than applicator (clinical margin)
6
Q
Anode angle
A
- the angle at which electrons hit the target
- effects focal spot size
- large angle = larger focal spot
- small angle = smaller focal spot
7
Q
Heel effect
A
- a loss in intensity on the anode side of the x-ray field caused by the attenuation of the x-ray beam by the anode
- refers to the variation in intensity across beam and to the cut off in the shadow behind the anode
8
Q
Imaging anode angle (4)
A
- want to avoid a blurry image = small focal spot required (reducing object to film distance also helps)
- area electrons hit on target must be small
- steep angle must be used to achieve this
- smaller anode angle (17 degrees)
9
Q
Imaging target type (4)
A
- imagine needs to be as fast as possible
- target exposed to large amount of electrons in a short period of time
- enormous amount of heat will occur on a very small area of target (very bad)
- target/anode must rotate so heat can be better distributed (very good)
10
Q
Therapy anode angle (4)
A
- we want a homogenous field = larger focal spot required (don’t need a clear image)
- area electrons hit on target must be larger
- flatter angle must be used to achieve this
- larger anode angle (45 degrees)
11
Q
Therapy target type (4)
A
- therapy can be delivered over longer periods of time (stabilizing set up)
- target exposed to large amount of electrons over a longer period of time
- heat is better distributed over target
- target/anode is stationary
12
Q
Anode cooling (3)
A
- x-ray tube requires time between uses to cool down
- amount of time it takes to set patient up is usually enough time
- x-ray tube also has protective mechanisms such as overload protection and water cooling which prevent damage