kV/MV FPI Flashcards
Isocentre check (quick)
Standard: +/- 2 mm
High precision: +/- 1 mm
FPI calibration check
No uncorrected defective pixels or gross artefacts
General imaging performance
Within specification
Imaging scaling
+/- 1 mm over 100 mm
Image centre alignment
+/- 2 mm
System ‘flex map’ or ‘IsoCal’ calibration
Standard: +/- 2 mm
High precision: +/- 1 mm
Isocentre coincidence (detailed)
Standard: +/- 2 mm
High precision: +/- 1 mm
X-ray beam collimation- variable collimators
+/- 2 mm over 100 mm
Ghosting
< +5% of baseline
Geometric distortion
+/- 2 mm over detector
Detector rotation
Within specification
X-ray beam collimation- fixed collimators
+/- 2 mm of baseline
Source and detector position
+/- 2 mm
Radiographic tests including dosimetry measurements
Within specification