What are common items in a HDR emergency kit?
forceps, stopwatch, pen, paper, digital dose monitoring devices, survey meter
what are the key assumptions made when performing shielding calculations for a HDR brachytherapy suite?
- patient doesn’t attenuate the radiation
- isotopic, all surfaces are considered primary barriers. Hence use factor (U)==1
- position of the source/patient in the room is fixed
what is used to determine the required attenuation of barriers for radiation safety: dose constraints or dose limits?
constraints, which are derived from dose limits (constraint<limit)
According to ICRP 103, what are the three fundamental principles of radiological protection? Which of these do not apply to therapeutic purposes?
Justification, optimisation, and dose limits. Dose limits do not apply for therapeutic exposures
According to ICRP 103, what are the three categories of radiation exposure?
Occupational, public, and medical exposures of patients.