KA 9 Flashcards

1
Q

What are common items in a HDR emergency kit?

A

forceps, stopwatch, pen, paper, digital dose monitoring devices, survey meter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

what are the key assumptions made when performing shielding calculations for a HDR brachytherapy suite?

A
  • patient doesn’t attenuate the radiation
  • isotopic, all surfaces are considered primary barriers. Hence use factor (U)==1
  • position of the source/patient in the room is fixed
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

what is used to determine the required attenuation of barriers for radiation safety: dose constraints or dose limits?

A

constraints, which are derived from dose limits (constraint<limit)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

According to ICRP 103, what are the three fundamental principles of radiological protection? Which of these do not apply to therapeutic purposes?

A

Justification, optimisation, and dose limits. Dose limits do not apply for therapeutic exposures

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

According to ICRP 103, what are the three categories of radiation exposure?

A

Occupational, public, and medical exposures of patients.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly