Instrumentation & Quality Assurance Flashcards
Which has an effect on the CRT appearance?
A) energy of isotope
B) dose calibrator
C) tag efficiency
D) count rate
D) count rate
As septal length _____, septal penetration _____.
A) increases, increases B) increases, decreases C) decreases, decreases D) decreases, increases E) both B and D
E) both B and D
The Z component of the signal in the photomultiplier tube will:
A) be processed by the pulse height analyzer
B) be recorded on the CRT
C) be removed by the PMT
D) all of the above
A) be processed by the pulse height analyzer
Image exposure time may be determined by:
A) a present number of counts
B) a preset time
C) information density
D) all of the above
D) all of the above
For a system with inverse gain control, and a gain setting of 4, each division on the pulse height analyzer dials corresponds to:
A) 0.5 keV/division
B) 1.0 keV/ division
C) 2.0 keV/division
D) none of the above
A) 0.5 keV/division
For a system with true gain control, and a gain setting of 4, each division on the pulse height analyzer dial corresponds to:
A) 0.5 keV/division
B) 1.0 keV/ division
C) 2.0 keV/division
C) 2.0 keV/division
What is the sensitivity of a camera that registers 28,137 cpm from a 120 uCi source is the background is 705 cpm?
A) 40 cpm/uCi B) 203 cpm/uCi C) 228 cpm/uCi D) 234 cpm/uCi E) 240 cpm/uCi
C) 228 cpm/uCi
How often should the COR check be performed?
A) at least weekly
B) before each pt.
C) every 6 months
D) at least monthly
A) at least weekly
The abbreviation COR stands for:
A) circle of rotation
B) center of region
C) correction of rotation
D) center of rotation
D) center of rotation
What is the minimum number of counts that should be obtained in a uniformity correction flood for a SPECT camera?
A) 10 K
B) 10 million
C) 30 million
D) 60 million
C) 30 million
Which of the following collimators will magnify an image?
A) flat field
B) diverging
C) converging
D) high resolution
C) converging
Which of the following must be performed every 6 months at minimum?
A) COR
B) sealed source leak test
C) linearity of the dose calibrator
D) area survey of hot lab
B) sealed source leak test
A dose calibrator must undergo repair or replacement if accuracy or constancy errors are:
A) present
B) greater than 5%
C) greater than 10%
D) grater than 20%
C) greater than 10%
Accuracy of the dose calibrator must be tested:
A) on installation
B) at least annually
C) every 6 months
D) both A & B
D) both A & B
Linearity of the dose calibrator must be checked:
A) on installation
B) quarterly
C) every 6 months
D) both A & B
D) both A & B