Instrumentation QC Flashcards
Energy Peaking is performed:
Daily
Uniformity test is done:
Daily
Linearity and resolutions done:
Weekly
Sensitivity test is done:
Weekly
T/F:
Energy peaking may be done with or without collimator
True.
Point source 5 UFOV from uncollimated camera
Sheet source may be used with collimated camera
Uniformity test
Point source 5 UFOV from uncollimated camera
Sheet source with C-57
Extrinsic uniformity test is performed:
Weekly/Daily
Daily. (With collimator, sheet source)
Intrinsic uniformity test is performed:
Weekly/Daily
Weekly. Point source 5 UFOV from camera
Extrinsic uniformity test:
Sheet source collimated camera
3-5 million counts
256 x 256 matrix
Visual and quantitative assessment
Integral uniformity:
Global or regional
Global
100% x [(max-min)/max+min)]
If max=min, then 0%
Lower is more uniform (
Differential uniformity:
Global or regional
Regional
Measure uniformity over a smaller area
100% x largest deviation (max-min)/(max+min)
Lower is more uniform (
Causes of non uniformity:
Cracked crystal
Malfunction PMT
Off peak or wrong peak
Resolution and linearity
Document spatial resolution
Evaluate ability to image straight lines
Flood source imaged through spatial resolution test phantom
Sensitivity test
counts/pixel/mCi
Time to acquire a press amount of counts using a set dose
Effect of nonuniformity
Ring artifact