Imaging in RT Flashcards
What factors should you consider when optimising imaging doses for RT?
1) For IGRT, produce images with a) dose ALARA and b) adequate image quality for the task
2) Higher doses may, utimately, result in lower doses overall for healthy tissue as improved image quality results in more accurate target localisation, thus sparing healthy tissue from additional (unintended) dose
3) 4D CT - for planning, can be justified if a) produces better quality CT images of moving targets b) allows more accurate measurements of target motion
4) Dose burden dose not change over time
5) Consistent image quality
6) Consistent performance of clinical tasks by trained operators
What are the factors that affect image quality and patient dose?
1) Image volume (FOV may be significantly larger than irradiated volume)
a) Detectors energy absorption efficiency
2) Irradiated volume
i. e. image to irradiated volume ratio
What does AAPM Report 180 contain?
Tables of organ doses from imaging and guidance.
What does AAPM Report TG132 contain?
Information regarding image registration (including evaluation).
What does AAPM Report TG179 contain?
Information on imaging dose for various modalities along with clinical applications, QA advice and some information on registration.
What factors should you consider when treating a bariatric patient?
1) Can the patient fit into the FOV of the scanner? Typical FOV for CT is 50-60cms; scanner bore 80-85cms.
2) Can the patient external be viewed int he FOV (not just the treatment area)? May need to use smaller arcs if not.
3) Extended FOV - artefacts at extents of field may mean the patient external not clear
4) “CT-ladders” used to determine positioning (metal clips on skin).
5) Does the patient’s weight effect couch movement and is it within the max weight limit of the couch? Limit approx 200-250kgs.
6) Reproducibility of positioning
7) placement of arms - needs for be able to place outside of treatment beam and same position when planning image taken.
8) if large 5-point mask needed, will this fit onto the couch?
9) increased patient size means more patient scatter, signal quality degraded, therefore can increase output from CT scanner, greater current, may have issues with beam on time (tube over-heating); also have max current therefore limit to image quality. Potentially increase slice thickness, reduced resolution in slice direction and could have issues with board placement during treatment. Could also use diagnostic scans (shorter scan) but not in treatment position.
10) may need to overwrite densities (e.g. CT-ladder)
11) Verification protocols - may need to change frequency from weekly to daily as more variation day to day
12) PTV margin increase may be necessary/change tolerance when setting up the patient for treatment
What are the requirements for a CT scanner for RT?
1) hard, flat couch top for reproducible positioning
2) compatibility with immobilisation devices
3) lasers for patient positioning
4) high geometric accuracy
5) Accurate map of HU to mass/electron density curves (RMI phantom) for each scan protocol used (as different energies therefore different linear attenuation coeffs); note for pediatric patients (due to different energies used, smaller body) at high mass/electron densities, actual values can deviate from the reference curve.
RT adult head - 120kV (predominant energy around 45kV)
Paediatric head - 80kV (predominant energy 25kV)
What is the equation to calculate HU?
1000 x [u - u(water)]/[uwater - uair]
From the literature, what are the recommended tolerances for the determination of electron density for:
a) Water
b) lung
c) bone
a) +/- 0.03
b) +/- 0.05
c) +/- 0.08
What are typical slice thicknesses for:
a) PET images
b) Planning CT
a) 4mm
b) 1.2mm
Why is PET imaging used in RT?
1) to determine lesion characteristics - stageing, metabolic rate, etc
2) assessment to treatment response
3) for trials - don’t need to store image, use generated metrics (lower IT storage requirements)
4) dose optimisation
5) testing drug targeting
6) RT target delineation (functional target volumes, therefore, aiding in reducing inter-observer variability compared with just CT alone)
7) PET can distinguish between tumour (metabolically active) and non-tumour (relatively metabolically inactive) masses, e.g. atelectasis in the lung.
Note: may have to take concomitant dose from PET imaging into consideration for RT plan.
What is the attenuation correction?
This corrects for the attenuation due to the patient from a coincident event (C) where the two annihilation photons are simultaneously detected.
C = C0 x exp[-uD]
C0 = attenuated signal detected D = total thickness through body/tissue u = linear attenuation coefficient
What is SUV?
Standard Uptake Value attempts to normalise uptake in tissues to the patient’s body weight.
SUV = activity concentration (Bq/ml) / [injected dose (Bq) / Body weight (g)]
Note an additional factor of: Glucose level in plasma (mmol/litre)/5.0 (mmol/litre), can be added to compensate for the uptake of glucose rather than FDG.
PET, SUV: what options could be used for body mass/weight?
1) Body mass/weight - most common as easy to calcuation but can be influenced by body shape
2) Lean body mass - more complex to calculate and there are different measurement/estimation techniques but lean body mass is more consistent across a range of body ‘habitus’ (body shape) for FDG; FDG very low uptake in fat as not as metabolic as muscle.
3) Body surface area - tallies with some dosing regimes
What are the three types of SUV values used for PET and what are their characteristics?
1) SUVmax: max value within an ROI, most commonly used as easy to determine but is dependent on the ROI drawn (inter-observer variation) and noise in the image can skew interpretation as making judgements based on a single pixel
2) SUVmean: mean value within an ROI; depends on ROI, therefore inter-observer variability but is less susceptible to noise than SUVmax
3) SUVpeak: uses small, fixed volume for an ROI and calculates the highest value; therefore, regional values do not affect it so it represents local higher values; this is a compromise between the other two types; ROI placement still varies with clinician but not volume; reduced noise influence; ROI volume and shape needs to be standardised if being used for comparisons.