image formation Flashcards
Which of the following statements is FALSE regarding the use of a narrow receiver bandwidth?
A. It generates lower signal-to-noise ratio images
B. It should be applied in anatomical regions that contain high fat and water interfaces
C. It is utilized with a high readout gradient amplitude
D. All of the above
All of the above
: A ______________ is defined as a picture element and has _________ characteristics.
Pixel, two-dimensional
The maximum intensity projection algorithm
Projects the high intensity pixels which represent blood vessels/ducts into one image
Two gradients applied at the same time during slice selection are utilized for
Encoding oblique slice planes
The readout gradient is usually turned on during the sampling or readout of the peak echo and also during which other process?
Frequency encoding
If the maximum slices in a set TR is 10, and the necessary number of slices for a given sequence’s anatomical coverage is 26, how many acquisitions (packages) will be required
3
Slice ___________ is determined by the slice selection process.
A. Position
B. Orientation
C. Thickness
D. All of the above
All of the above
What effect would decreasing the FOV have on the spatial resolution of an image
Increase
What effect would using a steep slice select slope and/or narrow bandwidth have on slice thickness?
Slices will be thin
Which set of scan parameters below would yield the best spatial resolution?
A. 270 mm FOV, 192 x 512 matrix, 5mm slice thickness, 8 NEX
B. 270 mm FOV, 256 x 512 matrix, 5mm slice thickness, 4 NEX
C. 200 mm FOV, 512 x 512 matrix, 3mm slice thickness, 4 NEX
D. 200 mm FOV, 512 x 512 matrix, 5mm slice thickness, 4 NEX
200 mm FOV, 512 x 512 matrix, 3mm slice thickness, 4 NEX
An isotropic voxel refers to a __________ shaped voxel.
Cubic
: If the radiologist requires 3mm slices axially acquired through the IAC, with a slice gap of 0.5mm, and requires 4.2cm total coverage, how many slices must be selected?
12
Anatomic slice coverage: Slice thickness + interslice gap x # of slices
If 24 4mm slices are planned with a gap of 1mm, the total anatomic coverage is _______ cm.
12
24 slices x 5mm (4 thickness + 1 gap) = 120 mm; ÷ 10 to calculate the coverage in cm.
Which of the following is NOT a result of reducing the FOV?
Anatomical structures are displayed smaller in the image
Creating additional images in various planes from a 3D data set is accomplished by a technique known as:
Multi-Planar reconstruction (MPR
Which of the following can be considered an advantage of selecting a 3D acquisition as opposed to a 2D acquisition?
A. The slices have a greater SNR
B. No crosstalk
C. Longer repetition times
D. All of the above
E. A & B only
A & B only
Multiplanar reconstructions are typically utilized in 3D pulse sequences because
A. 2D slices are too thick to reformat
B. 2D slice groups are too long
C. There are gaps between 2D slices
D. A & C only
E. B & C only
A & C only
The process that digitizes the MR signals is known as
Fourier transform
The signal produced immediately following an RF pulse is:
FID
__________ is defined as a small change in the magnetic field along a particular axis
A gradient
_________ is a file on the computer where the collected echoes are stored prior to being processed into an image by the Fourier Transform
Raw data
In order to generate an MR image, the slice select gradient must be turned on during
RF energy application