• Ice Bath Calibration Check Procedure Flashcards

1
Q

Prepare the ice bath. You need about 1/2 ice to water.

A

1

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

. Move the digital and air thermometers, as well as any meter probes, to the ice bath. Stir the ice
water continuously. You are looking for the digital thermometer to read as close to zero as
possible. If it is warmer, add more ice and mix.

A

2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Check the electronic meters again for stability and record the water temperatures to the
nearest 0.1 degree. Again, you will also record the NIST thermometer for comparison. Stir the
water to maintain even temperatures, adding ice if you move above zero on the digital
thermometer

A

3

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

. Now record the stick thermometer readings on the data sheet, to the nearest whole degree,
keeping the tips in the water while you take the readings. Make sure the data sheet is completely
filled in with your initials and the date.

A

4

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

. Notify staff of any instrument that doesn’t pass its QC test*

A

5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Put any unused ice back into the freezer.

A

6

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

. File your data sheet in the calibration notebook behind the “completed data sheets” tab, with
the most recent data sheet on top

A

7

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

*Electronic sensors must match within

A

0.2 degrees of the NIST

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

*Stick thermometers must match within

A

1 degree of NIS

How well did you know this?
1
Not at all
2
3
4
5
Perfectly