High Voltage Holiday Instruments Flashcards
High Voltage Pulse-type DC Holiday Detectors are designed to locate holidays in a 1___ coating applied over a 2___ substrate.
- Non-conductive.
2. Conductive
High Voltage Pulse-type DC Holiday Detectors are preferred for use on:
- ___ ___ coatings.
- ___ (fiberglass coatings) that could become electrostatically charged.
- Carbon impregnated.
2. Plastics
HV Pulse-type DC HD’s consist of:
- ___ ___ DC unit.
- Various probes for different ___ and ___.
- ___ accessories.
- ___ unit.
- __ wires.
- General ___.
- High voltage DC unit.
- Various probes for different substrates and configurations.
- Calibration accessories.
- Charging unit.
- Ground wires.
- General accessories.
Generally, each manufacturer/model has particular differences related to safety or operation that operators need to know.
For example: Tinker and Rasor’s APS Stick Type: is equipped with a release trigger that instantly turns the instrument off. Hold the handle down on the grip to operate the instrument. Release the handle and the unit shuts down. The feature increases job site safety since the instrument cannot be left energized while unattended.
The general operating procedure of most HV Pulse-type DC HD’s are as follows:
When the electrode is passed over the surface, a 1___ arcs through the air gap or coating to the substrate at any holidays, voids, or discontinuities, which simultaneously causes to detector to emit a sound.
The ground wire should be connected directly to the metal structure where possible. If impossible, use the HV HD with a 2___ ___ ___, provided the structure is also connected to the ground. If the building cannot be connected to the ground, establish direct contact or connect a ground wire at some point between the ground and the structure.
On concrete structures, attach the ground to rebar in the concrete, or lay the bare ground wire on concrete, cover it with a burlap bag filled with damp sand,.
Set the voltage as stated in either the specification or the referenced standard.
- Spark
2. Trailing ground wire.
If there is no referenced standard for the use of HV DC HD’s the rule of thumb is to use a voltage setting of ?
100 volts per mill of coating thickness.
When the voltage is set too high the coating may be damaged. The same damage can occur if the coating is tested before 1___ or ___ of the ___ has ___. Once a spark generates though the coating the substrate, a specific holiday now exists, even if there was none before the test was performed.
- Most or all of the solvent has evaporated.
When testing, move the electrode at a rate of 1__ ___ ___, in a 2___ ___ according to NACE SP0188.
- 1 ft per second.
2. Single pass.
Moving the prove too quickly may 1___ a ___. Moving too slowly may create 2___ at ___ ___, or prove to be more exacting than was intended by the specification.
- Miss a void.
2. Damage at thin spots,
Moving the prove too quickly may 1___ a ___. Moving too slowly may create 2___ at ___ ___, or prove to be more exacting than was intended by the specification.
- Miss a void.
2. Damage at thin spots,
Standards that may need to be consulted with use of HV Pulse type DC HD’s, depending on the specification and substrate type inlcude.
AS/NZS 3894.1
ASTM G62-07
ASTM D5162-08
ASTM D4748-08
NACE SP0274-2011
NACE SP0490-2007
NACE SP0188-2006
ANSI/AWWA C213-07
ANSI/AWWA C214-07
ISO 2746:1998
HV Pulse-Type DC HD’s come calibrated from the manufacturer and cannot be calibrated by the user. Check the 1___ ___ periodically, following manufacturer instructions. If there is ever a question about accuracy, take the detector 2___ of ___ and return it to the supplier for service.
- Voltage output.
2. Out of service.
Before using a HV Pulse-type DC HD ensure:
- Coating ___ is ___.
- __ ___ is complete.
- ___ of ___ is complete.
- ___ ___ is complete.
- Coating thickness needs to be above ___ ___.
- 2 LV testing of coatings below __ __ is complete.
- Coating cure is complete.
- Visual inspection is complete.
- Acceptance of inspection is complete.
- Thickness testing is complete.
- Coating thickness needs to be above 6 mils
- 2 LV testing of coatings below 6 mils is complete.
HV Pulse-type DC HD’s generally have a voltage output range from ___ to ___ volts.
800 to 60,000 volts.
HV Pulse-type DC HD’s are used on protective coating films ranging in thickness of?
6 to 240 mils.
When checking the output voltage from the probe, use of a high input ___ ___ is required to prevent the flow current and subsequent voltage drop from the power supply.
Impedance voltmeter