Hemodialysis Process, systems, and devices Flashcards

1
Q

How often are machine alarm tests performed?

A

Before the initiation of each patient treatment

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How long can a dialyzer be set up for once recirculation is complete? And why cant it remain set up longer?

A

2 hours
It can grow bacteria

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

On the Fresenius (FMS) dialysis delivery system, what should the manual conductivity value match on the display?

A

The conductivity value must be +/- 0.4 mS of the value displayed on the FSM

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What is the safe dialysate pH range?

A

6.9-7.6

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What is “strikethrough” of an external pressure transducer and why is it of concern?

A

It is when fluid could have entered the machine and contaminated the internal pressure transducer protector providing a reservoir for microorganisms and causing subsequent patient blood infections

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Why is it important for patient care teammates to know when the water treatment system disinfection was performed?

A

Because the end-to-end disinfection process will also introduce the disinfectant solution to the dialysis delivery systems through their water inlet lines.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly