handpiece safety Flashcards

1
Q

backcap

A

ensure cannot rotate

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

coupling

A

apply tension to ensure device secured

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

bur insertion

A

pull to ensure fully engaged

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

spin

A

check bur spins freely - roll on hand

should be no frictional resistance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

bur wobble

A

check bur lateral movement

should have a slight amount of lateral movement

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

final check

A

activate device for 5-10 secs to check sound

should be consistent pitch and volume

also look at the bur to check if it looks stationary (shouldn’t be blurry)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

if in doubt device fails any of these checks

A

notify dental nurse to get it checked

do not use on pt

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

check pack

A

expiration date of sterilisation

obvious fault s - dents, damaged to surface of optical rod etc

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

6 steps for check

A

3 safe to use
3 fit for use

  • Pinch outer rim of device head and apply rotation force, if backcap can be rotated (loose) send for repair
  • Pull bur to ensure fully engaged – apply significant tension
  • Fit device accessory in coupling – apply tension to ensure device secured
  • Check bur spins freely – frictional resistance between components may cause overheating during use (bearings need replacing)
  • Check burs lateral movement - too much deflection will cause bur to ‘wobble’ when activated, compromising control and cutting performance (bearings need replacing)
  • Check performance – run device for 5 sec and ensure it sounds right, and bur doesn’t wobble