Gas machine PPT Flashcards
Gas machines have a system to provide and measure (3):
- Oxygen
- Anesthetic agents
- CO2
SPDD (gas machine organization)
Supply
Processing
Delivery
Disposal
Supply for gas machine (2) - how do gasses get into the machine?
Pipeline and cylinders
O2 hospital pipeline liquid oxygen is converted to gas pressure of
50 psi
DISS stands for (and fucntion)
Diameter Index Safety alarm
non-interchangable, threaded fittings to connect gas powered devices to station outlets
cylinders on back of machine should be kept closed so that
alarm will sound with a loss of pressure
What is the most frequently missed item in preanesthesia setup
amubag
Cylinder valve
Most fragile part of cylinder
has a unique arrangement of holes that correspond with contents
Pipeline psi
50 psi
Oxygen cylinder is supplied at (psi)
1900-2200 PSI
Oxygen in cylinder is regulated to ____ PSI after it enters the machine
45 psi
NIOSH recommendation fo occupational exposure to nitrous
<2ppm halogenated agents (0.5 ppm if using N2O)
NIOSH reccomendation for occupation exposure to N2O alone
<25 ppm
What prevents one cylinder from emptying into the other
Free floating valve
E cylinder with highest pressure
closes valve on the cylinder with less pressure
Check valve in cylinder yoke (3):
- Minimizes trans-filling
- allows change of cylinders during use
- Minimizes leaks to atm if yoke is empty
Regulator in cylinder
supplies constant downstream pressure (45 psi) so that if cylinder is left open, gas will still flow
O2 service pressure
1900 psi
N2O service pressure
745-750
Air service pressure
1900
O2 capacity
660L
N2O capacity
1600 L
Air capacity
635 L
The five tasks of oxygen:
- Flows to the fresh gas flowmeter
- Powers the oxygen flush
- activates fail safe mechanics
- Activates oxygen low pressure alarm
- Compresses the bellows of mechanical ventilators
Flowmeters are located downstream from all safety devices and alarms EXCEPT:
the oxygen analyzer