Final Exam Flashcards
A finer speckled pattern generally has a higher transmit frequency. True or false?
True
Attenuation is expressed as Db(cm – MHz) the dash and the denominator indicates?
The depth and frequency affect attenuation
Which of the following listed is not a scanned modality?
- Sequentiale generated 2-D imaging
- PW Doppler
- Sequential generated color flow imaging
- 3-D imaging
- PW Doppler
What change would occur to the duty factor if the PRP increased by a factor of three?
Duty factor would decrease by a factor of three
What generally limits the maximum power allowed in non-scan modalities?
Risk of thermal bio effects
List four non-scan modalities of ultrasound.
- CW
- PW
- A mode
- M Mode
Define sampling rate.
The frequency of which signals are detected or assessed
Define frame time.
The time required to produce an image
Define acoustic beam.
A single transmit event in a specific direction and the associated echoes
Define frame rate.
The reciprocal of frame time
Define display line.
The data displayed on the screen that corresponds to a single direction within the patient
Define pulse wave.
A transmit beam that is turned on and off to provide range ambiguity
Define image.
To receive data for a specific time period formatted as a picture
Define receive beam.
The returning echo
What is the highest achievable axial resolution if the spatial pulse length is 12 mm?
Axial resolution equals SPL divided by two. So 12 ÷ 2 = 6.
What are the two ways to achieve better axial resolution as related to SPL equation?
- Shorter wavelength
- Fewer cycles in the pulse
The minimum PRP is the reciprocal of the maximum PRF. True or false?
True
Calculate the frame time for the sequential imaging scenario depicted.
Frame time equals 13 µs divided by centimeters times imaging depth divided by number of lines times number of lines
What would increase the frame rate?
- Decrease depth
- Decrease width
- Zoom
- All the above
- All the above
What imaging technique is depicted by the following figure?
- Plane wave imaging
- Sequential
- Parallel processing ✅
Plain wave imaging is a scanned or non-scan modality?
Non-scanned
What frequency does a Pedof probe normally operate when used in cardiac imaging?
1.9 MHZ
What was the main limitation of a mechanical transducer?
no ability to vary the focus