Final 712 Flashcards
What does a test box measure?
The performance of a hearing device in a controlled simulated environment
When are test box measures done? (4)
Before the hearing aid leaves the manufacturer
Before fitting a device in the clinic
Troubleshooting a patient complaint
After a hearing aid has been returned from a manufacturer repair
What can a test box do? (3)
Produces sounds of specific SPL
Attenuates ambient noise
Minimizes reflected sounds reaching the HA
What are the two HA couplers and what type of HA’s use each?
HA1 uses customs and RICs
HA2 is only for BTEs
When do we use putty?
Helps to seal sound in to make sure measurements of the HA is accurate is EAA
Usually used with customs because they are not standardized (it would cover the sound bore and vent)
Basic test box procedure
- couple the HA to the test box (ensure it is oriented the correct way, the front mic should be facing the front speaker)
- Paly specific types of measurement signals through the test box speakers (pure tone, broadband, speech)
- record the HA response to the signals
- compare the final measurement to a standard
Which ANSI standard defines HA performance?
ANSI S3
Why do we use ANSI standards?
Used as a quality check to determine HA performance
Provides equal ground with which to tell all HA (standardized measurement)
What is IEC?
International electrotechnical commission
Europe’s equivalent to ANSI
Full on Gain (FOG)
Gain control of HA is set to maximum output, this will generate a curve
High frequency average (HFA)
Average gain or SPL in decibels at 1000, 1600 and 2500
OSPL 90 curve
Reflects how the HA responds to a 90 dB SPL input when the HA is set to max gain
OSPL is the maximum output level if the hearing aid measured with a 90 dB SPL input signal (how loud is the HA capable of being)
Why is 90 dB used for the OSPL 90 curve?
Nearly always enough to saturate the HA
Reference test gain
HFA gain for an input of 60 dB when set to reference test position
Acoustic gain
At each frequency, the result obtained by subtracting the input SPL from the SPL coupler
What went in compared to what went out
This will have two gain values
Attack time
Time between the abrupt increase from 55 to 90 dB SPL and the point where the levels have stabilized within 3 dB of the steady state value for 90 dB input
How quickly does compression start after the onset of a loud sound
Equivalent input noise level (EIN)
Measurement of the internal noise (comparison between no sound and 50 dB input)
Reference test setting of the gain control (RTS)
Reduces volume to a position that will not saturate the hearing for mid-level input signals
Battery drain is measured at RTS
EAA steps with verifit 2
Calibrate, position HA in test box with correct coupler, connect HA to manufacturer software and begin test settings, FOG, RTS/gain, return HA to user settings, analyze measurements according to published spec sheet and ANSI tolerances
How do you calibrate the test box?
Under the test box test options, select calibration
No couplers are attached during this
Why do we need to calibrate?
Need to make sure that all measurements run in the test box are standardized and can be compared to each other and ensures that equipment is working
Positioning the HA in the test box with the correct coupler
EAA is always completed using blue coupler set
TRIC adapter attaches to HA1 coupler, otherwise would have to use putty- receiver should be pulled through but maintaining a tight seal
HA is centered over the “x” with front mic of the HA facing the front test box speaker
Reference mic is bent around to point at the front mic of HA
All HA run on the blue left coupler mic
How to run FOG?
Connect HA to manufacturer software using correct programming device and locate “test settings”
Some software do not have test settings, requiring the audiologist to put the HA into FOG and RTS manually (turn off all automatic features like feedback and noise management then max out programming bands for FOG, adjust gain down for RTS
On the verifit, under test box tests can select ANSI/IEC and follow directions
How to run RTS/gain?
Once FOG measurements are complete a green triangle will appear with the directions to “set VC to RTS”
The triangle with all but disappear once the HA has been programmed correctly