Final Flashcards
Relative Transmittance (T) definition
Fraction of incident light absorbed by the solution
Used as an index of analyte concentration
Relative Transmittance Equation
T = P/P0
P: radiant power of beam exiting cell
P0: radiant power of beam incident on absorption cell
Beer’s Law definition
Absorbance is directly proportional to concentration
Beer’s Law equation
A = abc
a: absorptivity
b: pathlength through solution
c: concentration of absorbing species
A = abc = ?
log(P/P0) = -logT = 2-log(%T)
Background Correction definition
processes other than analyte absorption (like reflection/scattering) result in significant decrease in power of incident beam
How to correct background?
Reference cell prepared by adding distilled water/solvent to an absorption cell
Placed in path of light beam, and the power of the radiation exiting the reference cell is measured and taken as P0 for sample cell
Absorbance equation using Background Correction:
A = log(Psolvent/Panalyte solution)
Psolvent: radiant power of beam exiting cell containing solvent (blank)
Deviation from Beer’s Law: (3)
- Only valid for diluted solution, up to mM (higher concentration, higher intermolecular distances, affect absorption of proton)
- Chemical processes such as the reversible association-dissociation of analyte molecules, or the ionization of a weak acid in an unbuffered solvent
- Instrumentation limitation (strictly applied to monochromatic radiation)
Reference solution in UV/Vis
Only solvent; chemically modify reference solution if sample compound is modified
Sample holder/cuvette in UV/Vis
Material doesn’t absorb any radiation in spectral region
Ex: quartz, silicate glass, plastic
How to choose appropriate wavelength in UV/Vis?
Choose wavelength at which the analyte has max absorbance and where the absorbance doesn’t change rapidly with changes in wavelength
How to Calibrate UV/Vis (0% and 100% Transmittance)
0%: shutter closed, set base level current, set 0%T
100%: shutter open, reference solution, set 100%T
Why would there be a nonlinear calibration curve in UV/Vis?
Due to concentration-dependent chemical change in the system or limitation of the instrument
Light Source in UV/Vis: (2)
Visible: tungsten filament lamp
UV: deuterium electrical discharge lamps; used with quartz holders,
Parts of UV/Vis spectrometer (5)
- Light source
- Monochromator
- Sample/reference holder
- Radiation detector
- Readout device
UV/Vis: Monochromator
To isolate the specific, narrow, continuous group of wavelengths to be used in the assay
UV/Vis: Radiation Detector
Produce an electric signal whn struck by photons
Signal is proportional to radiant power
UV/Vis: Readout Device
Analog meter or digital display showing transmittance or absorbance
Fluorescence Spectroscopy
Absorb energy from radiation in the UV/Vis range, then radiation simultaneously emitted when analyte relaxes
More sensitive than absorption spectroscopy
Key components of Fluorescence Spectroscopy (5):
- Light source
- Monochromator (emission and excitation)
- Sample/reference holder
- Radiation detector
- Readout device
Radiant power in fluorescence spectroscopy
Radiant power of the fluorescence beam emitted from a sample is proportional to the change in the radiant power of the source beam as it passed through the sample cell
Radiant power in Fluorescence equation
Pf = φ(P0-P)
Pf: radiant power
φ: quantum efficiency
Beer’s Law and Radiant Power
A = kP0c
(linear range; affected by pH, temp, solvent, impurity)