Exposure And Tecnique Errors Flashcards

1
Q

Film exposed to light
Appearance
Cause
Correction

A
  • Image appears black
  • Film was accidentally exposed to white light
  • don’t unwrap under white light, check darkroom for white leaks. Unwrap the film in darkroom
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Overexposed receptor
Appearance
Cause
Correction

A
  • dark
  • excessive exposure time from radiation, kilovoltage, or milliamperage
  • check exposure time
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Underexposed receptor
Appearance
Cause
Correction

A
  • light
  • inadequate exposure time kilovoltage or milliamperage
  • check receptor time, increase exposure time
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Incorrect receptor placement
Appearance
Cause
Correction

A
  • no apices are seen
  • not positioned correctly (Errors occur in the bisecting technique/paralleling)
  • make sure 1/8th of an inch of receptor edge extends beyond the incisal-occlusal surfaces of the teeth
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Incorrect horizontal angulation
Appearance
Cause
Correction

A
  • overlap contacts of the teeth
  • central Ray wasn’t directed through the Interproximal spaces
  • direct the x Ray beam through the Interproximal regions
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Incorrect vertical angulation
Appearance
Cause
Correction

A
  • teeth appear short with blunted roots on the image
  • was excessive(too steep) foreshortening occurs or elongation insufficient
  • to not excessive, use adequate vertical angulation
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Cone cut with beam alignment device
Appearance
Cause
Correction

A
  • A clear (unexposed) area is seen
  • PID was not properly aligned with periapical beam alignment
  • X-Ray beam is centered over the receptor
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Unexposed receptor:
Appearance
Cause
Correction

A
  • Clear
  • Failure to turn on x-Ray machine, electrical failure, machine malfunction