Exposure And Tecnique Errors Flashcards
1
Q
Film exposed to light
Appearance
Cause
Correction
A
- Image appears black
- Film was accidentally exposed to white light
- don’t unwrap under white light, check darkroom for white leaks. Unwrap the film in darkroom
2
Q
Overexposed receptor
Appearance
Cause
Correction
A
- dark
- excessive exposure time from radiation, kilovoltage, or milliamperage
- check exposure time
3
Q
Underexposed receptor
Appearance
Cause
Correction
A
- light
- inadequate exposure time kilovoltage or milliamperage
- check receptor time, increase exposure time
4
Q
Incorrect receptor placement
Appearance
Cause
Correction
A
- no apices are seen
- not positioned correctly (Errors occur in the bisecting technique/paralleling)
- make sure 1/8th of an inch of receptor edge extends beyond the incisal-occlusal surfaces of the teeth
5
Q
Incorrect horizontal angulation
Appearance
Cause
Correction
A
- overlap contacts of the teeth
- central Ray wasn’t directed through the Interproximal spaces
- direct the x Ray beam through the Interproximal regions
6
Q
Incorrect vertical angulation
Appearance
Cause
Correction
A
- teeth appear short with blunted roots on the image
- was excessive(too steep) foreshortening occurs or elongation insufficient
- to not excessive, use adequate vertical angulation
7
Q
Cone cut with beam alignment device
Appearance
Cause
Correction
A
- A clear (unexposed) area is seen
- PID was not properly aligned with periapical beam alignment
- X-Ray beam is centered over the receptor
8
Q
Unexposed receptor:
Appearance
Cause
Correction
A
- Clear
- Failure to turn on x-Ray machine, electrical failure, machine malfunction
- x Ray machine is turned on listen for the audible exposure signal