Equipment operation and quality control Flashcards
During Bremsstrahlung
(Brems) radiation production
An electron approaching a positive nuclear
charge changes direction and loses energy.
Design characteristics of x-ray tube targets that
determine heat capacity include
- the rotation of the anode
- the diameter of the anode
- the size of the Focal spot
If exposure Factors of 85 kV, 400mA, and 12ms yield an output exposure of 150 mR, what is the milliroentgens per milliampere-seconds(mR/mAs)?
(A) 0.32
(B) 3.1
(C) 17.6
(D) 31
If the output or 4.8mAs was 150mR, then 1mAs is equal to 31.25mR (150mR ÷ 4.8mAs =31.25mR/mAs).
If the output or 4.8mAs was 150mR, then 1mAs is equal to 31.25mR (150mR ÷ 4.8mAs =31.25mR/mAs).
Vs/Vp=Ns/Np
To determine secondary I
Ns/Np=Ip/Is
When using the smaller field in a dual- field image
intensifier?
the image is magnified
Light-sensitive AEC devices are known as?
phototimers
AEC devices has 2 types
1 type = ionization chamber
2 type = phototimer
AEC devices = ionization chamber function as
Beneath the table top above the IR, for centering the par centered to it sensor and radiographed.
When a predetermined quantity o ionization
has occurred , the exposure terminates.
AEC devices = phototimer
Small fluorescent screen, is positioned beneath the cassette. When remnant radiation emerging from the patient exposes the IR and
exits, the fluorescent screen emits light. Once a
predetermined amount o fluorescent light has
been “seen” by the photocell sensor, the exposure is
terminated automatically.
The low product of milliampere-seconds and kilovoltage determines in single phase anode
heat units.
The filament of the cathode
assembly is made of
thoriated tungsten
System function to
compensate or changing patient/part thicknesses
during fluoroscopic procedures?
Automatic brightness control
Increased total brightness gain does not affect
spatial resolution
Delivery of large exposures to a cold anode for the use of exposures exceeding tube limitation can result in
-cracking to the anode
-rotor-bearing damage
The high-voltage, or step-up, transformer functions to
increase voltage to the necessary kilovoltage.
Circuit devices that permit electrons to flow in
only one direction are
solid-state diodes
The slit camera is considered the standard for (annual) measurement of the
-effective focal-spot size.
-spatial resolution
It cause pitting, for many small surface melts, of the anode’s focal track?
Repeated, frequent overloading
There are three types of beam restrictors—aperture diaphragms.
cones, cylinders, and collimators.
The most practical and efficient type is the collimator.
The type of x-ray tube designed to turn on and off rapidly, providing multiple short, precise exposures, is
grid-controlled
T e device that receives the remnant beam, converts it into light, and then increases the brightness of that light is the
image intensifier
A high-speed electron is decelerated as is attracted to a tungsten atom nucleus. T is results in
Bremsstrahlung radiation
If the positive nucleus of a tungsten atom attracts the electron, changing its course, a certain amount of energy is released during the “braking” action. T is energy is given up in the form of an x-ray photon called
Bremsstrahlung (“braking”) radiation.
Is produced at the target when an incident electron ejects a K-shell electron, and an L-shell electron drops into its place. Energy is liberated in the form of
Characteristic ray, and its energy is representative of the deference in energy levels.
The rheostat is a type of variable resistor that is used to It is
change voltage or current values.
The procedure whose basic operation involves
reciprocal motion of the x-ray tube and IR is
tomography
In fluoroscopy, the automatic brightness control is used to adjust the
-kilovoltage (kV)
-milliamperage (mA)
When the radiographer selects kilovoltage on the control panel, which device is adjusted?
Autotransformer
The filament circuit supplies the proper
current and voltage to the x-ray tube for proper thermionic emission.
The rectifier circuit is
responsible for changing
AC to unidirectional
current.
Star and wye configurations are related to
three-phase transformers
The terms star and wye (or delta) refer to the
configuration of transformer windings in
three phase equipment.
high-voltage transformer windings has
three primary and three secondary windings—one winding or each phase
Autotransformers operate on
the principle of
self-induction and have only one winding.
What is the device that directs the light emitted from the image intensifier to various viewing and imaging apparatus
Beam splitter or objective lens located between the output phosphor and the TV camera tube (or CCD).
Characteristics of x-ray photons include
- Fluorescent effect on certain phosphors
- physiological effect on living tissue
Properties of x-ray photons are:
- X-rays are not perceptible by the senses.
- X-rays travel in straight lines.
- X-rays travel at the speed of light.
- X-rays are electrically neutral.
- X-rays have a penetrating effect on all matter.
- X-rays have a physiological effect on living tissue.
- X-rays have an ionizing effect on air.
- X-rays have a photographic effect on film
emulsion. - X-rays produce fluorescence in certain phosphors.
- X-rays cannot be focused.
- X-rays have a spectrum of energies.
- X-rays are unaffected by a magnetic field.
Congruence of the x-ray beam with the light field is tested using
radiopaque objects
Disadvantages of moving grids over stationary
grids include which of the following?
They increase patient radiation dose.
They can prohibit the use of very short
exposure times.
The functions of a picture archiving and communication system (PACS) include
reception of digital images
storage of digital images
The kV settings on radiographic equipment must
be tested annually and must be accurate to within
+ /–4 kV
The device used to test the accuracy of the x-ray
timer on single phase is the
spinning top
The x-ray tube in a CT imaging system is most
likely to be associated with
a pulsed x-ray beam
Single phase full-wave-rectified symbol.
(1ϕ )
Exposures less than the minimum response time of an AEC may be required when
using high milliamperage
using fast imaging systems
Which information is necessary to determine the maximum safe kilovoltage using the appropriate x-ray tube rating chart?
Milliamperage and exposure time
Focal-spot size
The speed of the imaging system has no impact on the use of a
radiographic rating chart.
A penetrometer can be used to demonstrate effects on
kV on contrast.
AEC devices (phototimers and ionization chambers)
automatically terminate the x-ray exposure once a predetermined quantity of x-rays has penetrated the patient, thus ensuring consistent results.
Defines the remnant x-ray beam
activates the phosphor within the intensifying screens, exposing the radiography film
The device used to ensure reproducible radiographs, regardless of tissue density variations, is the
automatic exposure control
Rectifiers are solid-state diodes made of semi conductive materials such as
-silicon
-selenium
-germanium that conduct electricity in only one direction.
To maintain image clarity in an image-intensifier
system, the path of electron flow from the photocathode to the output phosphor is controlled by
electrostatic lenses
Which of the following equipment is mandatory
f or performance of a myelogram?
tilting x-ray table
Deposition of vaporized tungsten on the inner sur-face of the x-ray tube glass window
acts as additional filtration
The brightness level of the fluoroscopic image can vary with
milliamperage
kilovoltage
patient thickness
Bremsstrahlung radiation or “braking” x-ray is produce at
high-speed electron, accelerated
toward a tungsten atom, is attracted (and
“braked,” i.e., slowed down) by the positively
charged nucleus and therefore is defected from
its original course with a resulting loss o energy.
Brems radiation comprises
70% to 90% of the x-ray beam.
Characteristic radiation comprises
very little of the x-ray beam between
(10–30%)
Characteristic radiation is produced at
high-speed electron encounters the tungsten atom and ejects a K-shell electron, leaving a vacancy in the K shell. An electron from a shell above (e.g., the L shell) shells the vacancy and in doing so emits a K-characteristic ray.
That portion of the x-ray beam striking the IR
and representing image anatomy is referred to
the
Signal
A high SNR meaning
(more signal, less noise).
SNR increases as
mAs increases
A computed tomographic (CT) imaging system
has three component parts:
a gantry
a computer
operator console.
A technique chart should be prepared for each
AEC x-ray unit and should contain which of
the following information for each type of
examination?
Photocell(s) used
Optimum kilovoltage
Backup time
Moving the image intensifier closer to the patient during fluoroscopy
decreases the SID
decreases patient dose
improves image quality
Although the stated focal-spot size is measured
directly under the actual focal spot, focal-spot
size actually varies along the length of the x-ray
beam. At which portion of the x-ray beam is the
effective focal spot the smallest?
At the anode end
The effective focal spot is always
smaller than the actual focal spot.
The total brightness gain of an image intensifier is the product of
flux gain*minification gain
Associated with magnification fluoroscopy?
Higher patient dose than nonmagnifcation
fluoroscopy
Higher voltage to the focusing lenses
Image intensifier focal point closer to the input
phosphor
Advantages of battery-powered mobile x-ray units include their
ability to store a large quantity o energy
ability to store energy for extended periods of
time
Off - focus is
responsible for indistinct images outside the collimated field
Off - focus, for extra focal, radiation is produced
as
electrons strike metal surfaces other than the
focal track and produce x-rays that emerge with
the primary beam at a variety of angles.
Which is/are components of the
secondary, or high voltage, side of the x-ray circuit?
Rectification system
The secondary coil of the step-up transformer and the rectification system are located in the
high- voltage (secondary) side of the x-ray circuit.
The timer, autotransformer, and (prereading)
kilovoltage meter are all located in the
low-voltage circuit.
A three-phase timer can be tested for accuracy
using a synchronous spinning top. The resulting
image looks like a
solid arc, with the angle (in degrees) representative of the exposure time
Anode angle will have an effect on the
severity of the heel effect
focal-spot size
heat-load capacity
The minimum response time of an automatic
exposure control (AEC)
is its shortest possible exposure time
Function to increase the
milliamperage?
Increase in heat of the lament
Formula would the radiographer use to determine the total number of heat
units produced with a given exposure using three phase, six-pulse equipment
mA × time × kV × 1.2
Heat units (HU) for a single-phase x-ray unit are determined by using the formula
HU = mA × kV × time.
For three-phase and high- frequency x-ray
equipment are determined by using the formula
HU = mA × kV × time × 1.4
Images using a particular single-phase, full-wave rectified x-ray unit were made, using known correct exposures. A spinning-top test was per formed at 200 mA, 0.05 s, and 70 kV, and 8 dots were visualized.
The 0.05-s time station is inaccurate.
The spinning-top test is used to test
timer accuracy for rectifier operation.
Because single-phase, full-wave-rectified current
has 120 useful impulses per second, a 1-s exposure of the spinning top should demonstrate 120 dots.
Therefore, a 0.05-s exposure should demonstrate six dots. Anything more or less than this indicates that the time station needs calibration. If exactly one-half the expected number of dots appears, rectifier failure is suspected.
The smaller the Focal spot
the more limited the anode is with respect to the quantity of heat it can safely accept.
Target angle and effective focal spot are
directly proportional
Target angle and actual focal spot are
inversely
Combination that will offer the
greatest heat-loading capability?
low degree target angle, high mm actual focal spot
The image-intensifier tube’s input phosphor functions
receives the remnant radiation emerging from the patient and converts it into a fluorescent light image.
(kinetic energy to light)
Very close to the input phosphor, separated by a thin, transparent layer, is the
photocathode.
Would be appropriate IP front material(s)?
Magnesium
Bakelite
The collimator light field and the actual x-ray field—they must be congruent (i.e., match) to within
2% of the SID.
Reproducibility testing should specify that radiation output be consistent to within
+ /–5%.
Examples of equipment components that are tested annually are the
-focal-spot size
-linearity
-reproducibility
-filtration
-kV
-exposure time
Kilovoltage settings can most effectively be tested using an electronic kV meter; to meet required standards, the kV should be accurate to within
+ /–4 kV.
The photocathode is made of
Antimony and cesium compound.
The fluorescent light image strikes the photocathode and is converted to an
electron image that is focused by the electrostatic lenses to the output phosphor.
If collimation is inadequate
and a field size larger than the part is used
excessive scattered radiation from the body for tabletop can cause the AEC to terminate the exposure prematurely, resulting in an underexposed image.
Accurate operation of the AEC device depends on
-positioning of the object with respect to the
photocell
-beam restriction
The AEC automatically adjust the
exposure required for body parts with different thicknesses and densities
T e image intensifier’s input phosphor generally is composed of
cesium iodide crystals
A photostimulable phosphor plate is used with
CR
radiography exposure refers to the
amount of the hit produce inside the x ray tube
Hit is the majority of energy that is produce
on a radiography exposure
An incorrect relationship between the primary
beam and the center of a focused grid results in
grid cutoff
insufficient receptor exposure
Spinning-top test is used to
test
timer accuracy or rectifier function in threephase equipment.
QA was being performed on a three-phase,
full-wave-rectified x-ray unit. A synchronous
spinning-top test was performed using 300 mA,
60 ms, and 70 kV. A 22-degree arc is observed on the test image. Which o the following statements regarding these results is most correct?
The test results are satisfactory.
Three-phase, fullwave-rectified current would expose a
360-degree arc each second, a 60-ms (0.06-s) exposure should expose a 21.6-degree arc (360 degrees × 0.06 = 21.6 degrees).
Will most likely to produce the highest quality radiographic image?
high SNR