Equipment Flashcards

1
Q

Gas Chromatograph

A

Absorption type analyzer
Measures Ne, O2, N2, CO
Measures multiple gases via a separator column
Requires carrier gas- He or Ne
Not very fast- not breath to breath
Affected by moisture and CO2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Mass spectrometer

A

Measures mass of gases
Measures multiple gases at the same time
Very fast- breath to breath
Very expensive and large

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Gas analyzer troubleshooting

A

Add the gasses and if not 100% there is an error- adjust the gain until the total equals 100%.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Infrared absorption analyzer

A

Measures CO, CO2 & methane
Uses a “chopper” motor/blade
Very fast- breath to breath
Calibration with room air (low), a known concentration of CO or CO2 (high), the room air again.
If calibration is off adjust the gain
Affected by moisture and long sampling line (delay)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Helium Analyzer

A

AKA thermal conductivity analyzer
AKA Wheatstone bridge analyzer
Helium passes over and cools hot wires and changes the resistance
Can’t be uses around combustible gases
Calibrate with room air, then a known concentration (around 10%), then room air again

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Nitrogen Analyzer

A

Gas sample pumped into a chamber where it’s ionized
Ionized nitrogen emits light- photo tube detects light
Very fast response, breath-to-breath
Troubleshooting: pump malfunction, leaks or excess moisture (causes erratic readings or spikes of nitrogen)
Calibrate with 100% O2, then known N2 concentration

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

ABG Analyzers: severinghaus electrode

A

Measures PCO2
Nylon spacer + teflon membrane
CO2 passes through into electrode to cause reaction

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

ABG Analyzers: Clark electrode

A

Measures PO2
Most complicated and most cause for errors
Platinum cathode + silver chloride anode
Polypropylene membrane- rough surface that can trap blood particles
Requires polarizing voltage
Cleaned periodically with pumice

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

ABG Analyzers: pH electrodes

A

Consist of reference and measuring electrodes
Reference electrode has constant electrical potential
Introducing a sample causes an electrical potential (charge) to develop between the electrodes
The voltage changes proportional to the pH

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Calibrating ABG Analyzers

A

PH electrode: add a sample of known pH and compare
CO2- low=5%, high=10%
O2- low=0%, high=12% or 20%
To read results, need to convert % into mmHg (713 x % as decimal)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Troubleshooting ABG analyzers

A

Ensure gas tanks aren’t empty
Unstable, erratic or noisy readings: worn out or ruptured membrane, clots (protein remover + flushes to clear)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Galvanic fuel cell analyzer

A

AKA zirconium fuel cell
Expose to O2 to create reaction & analyze
Fast and accurate
Measures PO2 and converts to %
Affected by changes in pressure or altitude
Troubleshooting- only malfunction if fuel runs out- replace

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Polarographic O2 analyzer

A

Uses a Clark electrode to measure PO2 converted to FiO2
Affected by water on sensor, changes in pressure or altitude
Troubleshooting: check batteries, electrolytes, membrane

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Calibrating O2 sensors

A

FiO2: 21% & 100% O2
FeO2: 12-14% & 21% O2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Tonometer

A

Used to QC ABG analyzers
Creates a precise gas mixture in a buffering solution or blood
Adds known PO2 or PCO2 to QC blood sample to check analyzers

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Valves & sampling devices

A

Ensure proper function: observe the flaps or diaphragms aren’t sticking
Should have low dead space, low resistance, low compliance (stiff), easy to clean
To determine dead space- fill with water and drain out into a graduated beaker

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

CO2 absorber

A

AKA CO2 scrubber or sodium/barium hydroxide canister
Removes CO2 from air and converts to water
Only necessary for tests requiring rebreathing
Malfunctioning CO2 absorber can cause: flushing, dizziness, light-headedness, increased RR (stop testing and check system)
Color change= chemical used up and should be replaced (usually brow->white or cream->purple)

19
Q

Water absorber

A

AKA desiccant
Calcium sulfate (dry-rite)
Placed inline after CO2 scrubber
Blue -> pink = used up

20
Q

Perma Pure tubing

A

AKA Nafion tubing
Allows humidity in tube to equilibrate with room air

21
Q

Wright respirometer

A

Measures flow and estimates volume
Exhaled flow spins a vane and turns gears that then turn dial to show volume
Convenient but also delicate and not very accurate

22
Q

Pressure differential pneumotach

A

Measures a pressure differential and converts to volume
Contains resistive elements (typically capillary tubes)- flow over the elements changes the pressure on one end vs the other
Quick and accurate
Moisture in the capillary tubes can cause high volume errors- heating element helps minimize moisture
Troubleshooting- check for moisture, check or replace heating element

23
Q

Thermal element pneumotach

A

Heated wire cools as air passes and converts to flow
Contains a screen with heated wire behind it- damaged or blocked screen causes errors

24
Q

Pilot tube

A

Not affected by moisture
Contains two tubes to measure inspiratory and expiratory flow
Disposable

25
Ultrasonic pneumotach
Flow passes over a strut (obstruction) and creates vortexes/waves Ultrasonic transducer counts the waves to calculate flow Affected by moisture
26
Volume Collins water seal spirometer
Records volume via a moving bell that marks on a rotating drum Mechanical and very accurate
27
Calibration syringe
3L syringe that must be accurate within +/-3.5% (105mL/.15L) - 2.90-3.10L is good Too low = leak, too high = potentiometer needs adjustment If readings are consistently high/low, a correction factor may be used to compensate AKA super syringe or 3-literally syringe
28
Dry rolling seal spirometer
Dry piston moving side to side directly measuring volume Exhalation pushes piston away, inhalation pulls it back May be connected to a pen or a potentiometer
29
Wedge bellows spirometer
Inflates and deflates to directly measure volume Pen moves up and down with breathing A hole in the bellows creates a leak and reduced volumes
30
Rotometer
Calibration device that generates specific flows
31
Sine wave rotary pump
Pumps air in and out of a body box to generate pressure changes and calibrate
32
Body box
Based on Boyles’ law- relationship between pressure and volume Only works if temperature is constant Measures TGV- more accurate for calculating FRC than N2 washout or He dilution Also measures Raw Calibration: mouth pressure with barometer, box pressure with sine wave rotary pump Can compensate for small calibration errors by adjusting the gain
33
Calibrating treadmill
Measure belt length Count revolutions per minute Calculate speed
34
ECG tracing
1mV = 10mm vertical deflection Lead II = best limb lead. Upper R arm negative, L leg positive (upper right to lower left, following normal heart conduction) Limb leads = overview of heart conduction Chest leads (V1-6) = close evaluation of heart function V5 = best chest lead (sits over L ventricle)
35
Air-oxygen blender
AKA air-oxygen proportioner Alarm = air or O2 side has < 40psi pressure
36
Barometers
Mercury barometer Vacuum tube- pressure pushes against mercury Water barometer (aka u-tube barometer) Air pressure on one side pushes up on the other side
37
Inductive plethysmography
Measures chest movement with resistive bands
38
Aneroid manometer
Measures air pressure
39
Integration
Taking a flow signal and turning into volume
40
Chemical disinfection & sterilization
Ethylene oxide gas (ETO)- sterilize Safe for all equipment Alkaline gluteraldehyde (Cidex)- short time (20 min) = disinfection, long time (overnight) = sterilize Bleach- disinfect
41
Metabolic measurement systems for exercise testing
Mixing chamber- mixes gases before analysis Breath-to-breath- analyses gas in each exhaled breath