eLFH - Invasive monitoring Flashcards
Components of invasive arterial blood pressure monitoring
Arterial cannula - Hagan-Poiseuille
Fluid filled tubing - Damping and resonance
Transducer - Wheatstone bridge circuit
Signal processor - Fourier and pulse contour analysis
Amplifier - Amplification
Display - Calibration
Hagan-Poiseuille equation
How does arterial cannula determine flow
Hagan-Poiseuille equation
For arterial BP measurement:
- Pressure difference maintained by using pressure bag set to 300 mmHg
- Cannula radius is fixed (but clot in lumen may change this)
- Length is fixed
- Fluid viscosity is assumed to be constant
Mistake which could alter viscosity in invasive BP measurement system
Inadvertent use of 5% dextrose instead of normal saline
Dynamic response definition
Speed at which it is able to settle on a new value following a stimulus
Three factors which affect dynamic response of an arterial line system
Natural (resonant) frequency
Input frequency
Damping
Natural (resonant) frequency definition
Frequency at which a system oscillates when set in motion
Unique for each system
Represented by sine wave
E.g. tuning fork vibrates at its natural frequency
Input frequency definition
Frequency of energy input into the system
Resonance definition
Effect observed when input frequency is the same as natural frequency
If energy is input at same frequency as natural frequency, then amplitude of swing increases exponentially
If they don’t match, then amplitude decreases overall
E.g. playground swing increases amplitude if energy is input at same frequency as natural frequency
Damping definition
Energy loss of a swinging or oscillating body through friction / resistance
Damping representation
Damping represented by the Damping Coefficient (D)
Under damping definition
Takes very long time for system to settle on a new value (zero on x axis following stimulus)
Damping coefficient < 0.64
Critical damping definition
Damping coefficient = 1.0
Characterised by no overshoot and long time for amplitude to settle at zero
Optimal damping definition
Damping coefficient = 0.64
Takes shortest time for amplitude to settle at zero
Typically has 1x overshoot and then settles (i.e. over reads once, under reads once and then settled)
Graphical representation of dynamic response and relationship between input frequency, natural frequency and damping
Maximum response when input frequency : Natural frequency ratio = 1 (i.e. input frequency = natural frequency)
When does resonance occur in arterial line system
When input frequency (heart rate) = the natural frequency of the system
Effect of resonance on arterial line trace
Significantly reduces quality of arterial line trace
How to avoid resonance in arterial line (or other) system
Natural frequency of a system should be at least 8x greater than the maximum anticipated input frequency
Minimum natural frequency of a system in clinical use to avoid resonance
20 Hz
Method used to determine natural frequency and level of damping in an arterial line system
Flush test
Calculation of natural frequency of a system using flush test
Calculation of damping level of a system using flush test
Optimal damping coefficient for all clinical systems to provide best dynamic response
0.64
Transducer definition
Device that converts one form of energy into another
Arterial pressure transducer definition
Converts pressure energy into electrical energy
Three types of transducer commonly used in arterial pressure measurement
Wire strain gauge
Bonded strain gauge
Capacitive transducer