Echo Physics 101 Flashcards
Fundamental Imaging
Based on the reflection of transmitted frequency. US waves pass through tissue twice
Harmonic Imaging
harmonic frequency is generated as the US signal propagates through the tissue. It is single-pass imaging and therefore reduces artifacts. Useful for imaging deeper structures
Reverberation - define
how to fix
more distant to true object
comet/ring down
straight line through probe center
fix: decrease gain
alternate imaging plane
Acoustic shadowing
how to fix
alternative imaging
increase gain or adjust TGC
mirror artifact
more distant than true object
decrease gain
refraction artifact
at same distance from probe
decrease gain
use alternative imaging planes/avoid refracting structure
side lobe artifact
linear, symmetric at both sides of object
same direction from probe (arc-like in radial direction)
apply color doppler
decrease gain
beam width artifact
at same distance from probe
true object/doppler signal outside imaging plane
adjust focal zone
alternative imaging plane
near field clutter
apply color doppler, reducing scale
alternative imaging plane
doppler shift equation
[2xreflector speed x incident freq x cos(theta)]/propagation speed
doppler shift is directly related to _
blood cell speed
freq of transducer
cos of angle bet. flow and sound beam
doppler shift is inversely related to _
speed of sound in medium
what does the 2 in doppler equation represent
double doppler shift : 1st is when sound strikes cell, 2nd is from moving cell reflecting wave back to transducer
in order to accurately determine velocity…
the angle between direction of flow and sound beam must be known
velocity (measured) is as related to Doppler shift…
= true velocity x cos (theta)
Doppler shift definition
a change or variation in the frequency of sound as a result of motion bet. sound source and the receiver
Doppler frequency
difference between received and transmitted frequencies
Positive doppler shift
when source and receiver are approaching each other
reflected freq > transmitted freq
Negative doppler shift
when source and receiver are moving apart
reflected freq < transmitted freq
typical values for audible sound
20 Hz to 20 kHz
doppler US uses these transducers
2 to 10 MHz
Demodulation
extracts Doppler freq from transducer freq and is performed by a demodulator
bi-directional doppler is analyzed with
phase quadrature processing
doppler shift alternate equation
received - transmitted freq
relation bet. velocity and doppler shift
direct
duplex US
simultaneous imaging and Doppler
disadvantage of CW Doppler
range ambiguity - echoes arise from entire length of overlap between transmit and receive beams
range = depth
Nyquist limit
prf/2 (in kHz)
aliasing appears when
doppler shift exceeds the Nyquist limit
5 ways to Eliminate Aliasing
use CW
use lower freq transducer
(reduces doppler shift and shrinks spectrum)
select new view with shallower sample vol (increases PRF and Nyquist limit)
increase scale
baseline shift