Drager Modes & Variables Flashcards

1
Q

What are three extended configurations offered on Drager vents?

A

Pressure Support
Autoflow
Volume Guarantee

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

When a breath is initiated by the patient, it is called ___. They are ___ triggered breaths.
When it’s triggered by the vent, it’s called ___. They are ___ triggered breaths.

A

Patient triggered, flow cycled

Mechanically triggered, time cycled

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

According to Drager, when should alarm limits be checked?

A

With every patient admission and every change in ventilation mode.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

For a patient on VC, what’s the most important alarm(s)?

A

Phigh

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

For a patient on PC, what’s the most important alarm(s)?

A

VThigh
VTlow
MVhigh
MVlow

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

In volume control, the flow is ___.

A

Constant

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What would happen if the inspiratory flow is so high that the set volume is reached before the set I time has passed? (Not applicable if ___ is enabled.)

A

There will be a pause in inspiration.

Autoflow

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

There are four VC modes on the Drager. What are they?

A

VC-MMV (Mand. Minute Vol.)
VC-AC
VC-CMV (Cont. Mand. Vent)
VC-SIMV

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Autoflow is available in what mode?

A

All VC breaths

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What does Autoflow do in regards to pressure?

Flow curve will not be constant, it will be ___.

A

Ensures that the set tidal volume is applied with the necessary minimum pressure for all mandatory breaths.
Decelerating.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

On Drager graphics, blue waveforms mean the breath is ___ triggered. Gray breaths mean it’s ___ triggered.

A
Blue = Machine
Gray = Patient
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

In MMV, the patient can breathe spontaneously. If the spontaneous breathing is not sufficient to achieve the set MV, ___ triggered, ___ cycled mandatory breaths are applied. Pressure support (may/may not) be applied.

A

Machine, Time

PS may be applied

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

In VC-MMV, the set breathing frequency is the (minimum/maximum) number of mandatory breaths.

A

Maximum

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

There are six PC modes on the Drager. PC-CMV, PC-AC, and PC-SIMV are three of them. What are the other three?

A

PC-BiPAP (Biphasic Pos Airway Press)
PC-APRC
PV-PSV

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

In PC the flow pattern is ___.

A

Decelerating

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What does volume guarantee do?

A

Ensures that for all mandatory breaths, the set Tv is applied with minimum pressure.