Dose 1 Flashcards

1
Q

What is a subjective assessment

A

Purely personal and likely to be different to that of another individual

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Objective assessment

A

Refers to quantity that can be measures and assigned a values. Independent of personal opinion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

what are 3 means of objective assessment

A

EI, LEEDS test objects (contrast sensitivity), CDRAD (physical properties)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

sensitivity

A

probability of a positive result among those with conditions

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

specificity

A

Liklihood for negative result for those without condition

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

what is troubleshooting

A

A logical systematic search for the source of a problem in order to solve it

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

what are the 3 resources for troubleshooting

A

manufacturer resources (service engineer, applications specialist, equipment manual, user guide), equipment logbook, QC reports

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

6 steps to troubleshooting

A
  1. verify that problem actually exists
  2. articulate symptoms
  3. determine circumstances where it occurs
  4. brainstorm possible causes
  5. rank causes in probability
  6. investigate causes from most likely
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

what are the 2 types of quality measures in medical imaging

A

people centred quality (patient care), product centred quality (diagnostic images)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

what is the goal of QC testing

A

to detect change in equipment function from the original level of performance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

what is the QC objective

A

enable prompt corrective action to maintain image quality and radiation safety

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

what is the QC rationale

A

degredation in equipment performance is gradual and expected, testing uncovers changes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

what is the difference between acceptance and compliance testing

A

acceptance testing measures standards at installation, compliance testing is performed periodically to ensure continued compliance with standards

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

what are the benefits of QC testing

A

detection of equipment deterioration, ALARA, ensures patient and staff safety, ensures legislative compliance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

what are the four stages of checks

A
  1. critical examination- ensure safety features working
  2. acceptance- verify all equipment Is supplied
  3. commissioning- by engineer, ensures equip is safe for clinical use
  4. routine performance testing- reg testing throughout lifetime
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

what is the difference between a run chart and a trend chart

A

a run chart displays data in time sequence, a trend chart is simlar but has ideal central line with upper and lower limits

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

how do you interperate a trend or run chart

A

look for trends (atleast 3 sets of data in same direction) in same direction. when a first measurement is out of limits it needs to be remeasured, corrective action needs to be performed when a trend indicates that process is out of control

18
Q

what are the QC tests for the xray generator

A
  1. filtration/HVL test
  2. radiation output test
  3. reproducibility test
  4. reciprocity test
  5. linearity test
19
Q

what does the HVL test look for, when is it performed and what are the steps

A

ensures adequate filtration to low dose photons of beam, performed for new XRT and needs to meet manufacturer standard. Steps: collimation set to 0, dosimeter is used and exposures are taken with increasing layers of Al. percentage transmission is calculated as exposures are taken and is plotted against thickness, 50% transmission = ideal amount. cannot fall below 2.3mmAleq

20
Q

what does the radiation output test look for, when is it performed and what are the steps

A

it assesses the lines voltage compensation and microprocessers of the XRT, and annual testing is required. for indirect reading: rad system with internal filter- one image taken with and one image taken without filter- compares two readings and LUT used to determine what causes different readings. for direct read: output recorded by dosimeter. % accuracy calculation performed. needs to be within 10% of baseline

21
Q

what does the kV reproducibility test look for, when is it performed and what are the steps

A

measures the ability of the generator to deliver the same output when the same exposure factors are used. dosimeter placed In ray and 5 exposures are taken; reproducibility variance % calculated from the readings. variance must be below 5%

22
Q

what does the reciprocity test look for, when is it performed and what are the steps

A

it measures the ability of the generator to produce the same output given the same kVp and mAs, with alternating mA and time settings. performed anually. 5 exposures with same kVp and mAs (different combos) are made, dosimeter used to record output. reciprocity variance calculated for max and min measurements, must be below 10% variance

23
Q

what does the linearity test look for, when is it performed and what are the steps

A

testing of mA and timer separately to determine the proportionality of output with increasing exposure factors. completed annually. mA linearity steps; mA increased while all other factors kept constant. timer linearity test: time increased with all other factors kept constant. 5 exposures are made, reciprocity variance used and must be below 10%

24
Q

what does the xray beam/grid alignment test look for, when is it performed and what are the steps

A

tests for lateral decentering or tilting of the focussed grid in the bucky. performed annually. lead panel with 5 holes, each hole exposed covering others, middle hole should be most dense. decentering needs to be within 1/2 inch or +/- 2% SID. failed test is remedial

25
what is the xray field IR test, when is it performed and what are the steps
ensures that the x-ray field and IR coincide when XRT and bucky are aligned. performed when grid is replaced or XRT serviced. steps; 24x30 cassette placed in bucky, indicator lights green, 2x sets of exposures taken; one pulling the XRT toward you, one pushing it away. sum of variance of both sides must be less than 2% SID
26
what is the grid uniformity test, when is it performed and what are the steps
ensures grid working effectively, performed annually?, PSP cassette erased, cassette on floor and SID focal range of grid used, collimate to full cassette and expose with tib fib protocol, perform test again with grid on cassette, reprocess with mammography filter and view both images for uniformity- ROI of 5 corners must be within 10% variance
27
what are the 3 components that contribute toward image quality
1. psp cassette (captures latent image) 2. CR reader/digitizer (processes image and applies algorithm) 3. digital electronics (projects image)
28
what are the components of the CR reader
1. mechanical- rollers 2. optical- laser lights 3. computer control- laser tagging (processing of images)
29
when should spot checks of the CR cassette and IP be performed
daily or weekly with cassette chosen at random. inspection/cleaning performed weekly or monthly (disinfectant + microfibre cloth)
30
what is the background noise test, when should it be performed and what are the steps
analyses image for unwanted background noise, 18x24 cassette scanned under hand examination, brightness and contrast scaling performed, 5 ROI measurements performed. EI should be NULL
31
what is the ghosting test, when is it performed and what are the steps
analyses image for ghosting (superimposition of previous image over current image) caused by inability to deactivate detector plate phosphors that were energised on previous exposure. steps; take extremely overexposed phantom image, put psp through normal cycle, put psp through normal cycle again, take ROI and compare. failed test- IP needs to be replaced
32
what is the flat field test, when is it performed and what are the steps
ensuring that there is pixel uniformity in FFD, 35x43 cassette with open collimatioin, process plate with high contrast filter (mammography), view image for non uniformity and ROIs, must be within 10% variance
33
what are the 3 components of the DR system
1. image receptor 2. image detector 3. image processing
34
what is a del and what is the fill factor pixel ratio
the DR image receptor is a configuration of Dels, containing TFT, pixel sensing area and storage. fill factor of pixel is the ratio of sensing area of pixel to area of pixel itself. high fill factor is better
35
what are the 3 stages of detector calibration and why are they performed
offset correction, gain correction and pixel calibration
36
what is offset calibration and how is it performed
ensures that each pixel in the detector measures radiation from the same level. steps; logged in as DR -administrator, system-general-quality assurance, entries for detector calibration and follow screen prompts
37
what is gain calibration and what are the steps for it
ensures that all pixels in the detector generate the same signal value for the same amount of radiation. align xrt to wall stand with 150cm SID, turn key in XRT to manual override, open collimation to 2cm larger than sensing area with 0 filtration, insert 21mm Al filter and no grid, select gain calibration in patient list and perform 10 exposured
38
what is pixel calibration and what are the steps
addresses dead pixels and applies algorithms based on neighbouring pixel values. steps; select pixel calibration from patient list, perform the sequence of 12 exposures
39
what is the SMPTE task for monitor function
assesses if monitor is displaying all signal information w/o loss. steps; patient list- test images- review- spmte. select scaling tool and locate full screen and perform visual inspection. for contrast, black and white squares should be clear. for uniformity, background grey should look uniform. for spatial resolution each corner line pattern should be visible
40
what is the kVp compensation test and what are the steps for it
assesses for consistant level of image quality at different tube potentials. steps; 4 images of test phantom taken with all factors consistant. mean EI and ROI is calculated. EI and ROI of the images must not be more than 20% or less than 15% of average
41
what is the AEC reproducibility test and what are the steps
assesses for consistency of the AEC chambers in recording the signal received. 2 sets of images taken. test 1; phantom placed over central chamber and exposed 5 times, radiation meter records air kerma, mean Is determined of air kerma and EI. smallest and largest value is calculated and expressed as % of avg. test 2; phantom placed over left than right chamber and 2 images taken, air kerma and EI average calculated, smallest of 2 air kerma averages must be within 10% of all other values
42
what is the patient thickness compensation test
determines if AEC is adjusting appropriately for changing patient thickness. test phantoms of varying thicknesses exposed at different kV settings, EI and ROI recorded, average of data recorded. average EI must not differ by more than 20% of the lower two the EI recorded, average central signal value must not differ by more than 20% of the lower two signal values