Dorado Flashcards
What are Dorado balloons indicated for?
PTA of renal, iliac, femoral, popliteal, tibial, peroneal, and subclavian arteries
True or False: Dorado PTA can be used in coronary arteries
False
What % balloon growth does Dorado PTA have from nominal to RBP?
1.5%
What % balloon growth does Mustang PTA have from nominal to RBP?
8.62%
True or False: Mustang PTA balloons can have the “dog bone” effect when inflated
True
True or False: Dorado PTA balloons can have the “dog bone” effect when inflated
False
What type of balloon is a Dorado PTA?
ULTRA non-compliant
Using (30) balloons to test to failure, what % of Mustang PTA ruptured at the RBP?
50% failed at RBP
0% failure for Dorado
What is the range of diameter option for Dorado PTA?
3mm - 10mm
What are the balloon length options for Dorado PTA?
20mm - 200mm
What are the catheter shaft length options for Dorado PTA?
40cm
80cm
120cm
135cm
What is the nominal (atm) for Dorado PTA?
8atm
What is the recommended guidewire size to use with Dorado PTA?
0.035