Devices/EP Flashcards
Class 1 Indications
3RD DEGREE & ADVANCED SECOND DEGREE AVB:
➢ Symptomatic bradycardia due to AV Block
➢ Documented periods of asystole >3 seconds in duration
➢ Escape rate of <40 bpm while awake; symptom free patients
➢ Post AV junction ablation
➢ Post-OP AVB not expected to resolve
➢ 2nd degree AVB regardless of type or site with associated symptomatic bradycardia
Other indications
Vasovagal syncope (VVS)
Prevention and termination of tachyarrhythmias including prolonged QT syndrome (pause-dependant)
AV Block associated with MI’s
Hypertrophic & Dilated Cardiomyopathies
Veins used for pacing
subclavian
cephalic
internal/external jugular
What is Impedence?
Represents the opposition to current flow
If impedene increases, current flow will…
decrease
Impedence values
300-1000ohms
Benefits of dual chamber pacing
Provides AV synchrony
Lower incidences of atrial fibrillation
Lower risk of systemic embolism and stroke
Lower incidence of congestive heart failure
Higher survival rates
Increased LV filling and resting cardiac output
Magnet response
Prevents sensing, resulting in asynchronous pacemaker operation
allows assessment of pacemaker function
What MODE(S) would you typically see once a magnet is applied over the device?
VOO
DOO
AOO
Pacemaker Interogation
Programmed parameters
Measured data
ECG/EGM and marker channels/diagrams
Diagnostic information
Battery status
Lead status & impedance
Rate histograms
Sensing
P wave amplitude at time of implant should be atleast greater than 2mV; and the R wave amplitude >5mV
Oversensing =
Underpacing
Undersensing =
Overpacing
Causes of Loss of Capture
Lead dislodgement
Lead insulation break
Lead wire fracture
Electrical circuit failure
Battery depletion
Pacemaker or lead connector issues
Pacing/sensing parameters programmed too low or high
Electrolyte imbalance
Exit block at the lead fixation site (loss of capture)
Drug effects
Upper Rate Behaviour
the pacemakers response to high atrial rates
What mode for AFIB?
VVI - do not want to pace the atrium
OR
VVIR - rate responsive
ICD’s must be able to
Sense small and large R waves
NOT sense T waves
Sense very small VF signals
Ignore and not sense activity in opposite chambers
Charge time
The time between the detection of an arrhythmia and when the therapy is delivered.
Initiated upon the detection of an arrhythmia
Consequences of Undersensing
Failure to sense VF, therefore, no appropriate therapy will be initiated
If VT goes undetected, the failure to treat it can quickly deteriorate to VF
Consequences of Oversensing
Delivery of inappropriate therapies
This can lead to inducing fatal arrhythmia
ICD Therapies include
Anti-tachycardia pacing
Defibrillation
Cardioversion
Damage to insulation can cause
Low impedance
increase in current flow
accelerated battery depletion
Lead conductor fracture can cause
High impedance
decrease in current flow
artifact/noise on EGM
Pacemaker cells
SA Node
AV node
Purkinje fibres
P-A interval indicates
approx time taken from the SA node-AV node
A-H interval indicates
approx time it takes to travel over AVN
represents conduction time from low RA through AVN and bundle of HIS
H-V interval indicates
Approx time it takes for HIS-purkinje system to ventricular myocardium
Fast response cells
Cardiac myocytes
purkinje cells
Slow response cells
SA node
AV node