Device, algorithm & feature selection Flashcards
1
Q
In what patients would an additional coil in the azygous vein be appropriate?
A
Those with high shock impedance
Those with multiple failed shocks
2
Q
Why is ICD sensitivity more sensitive following a paced beat?
A
To prevent pacing into VF