Day 1 Flashcards
Definition of the Intox EC/IR II
An approved instrument which analyzes a sample of deep lung air and reports results in milligrams of alcohol in 100 millilitres of blood.
Henry’s Law applied to breath testing.
At 34 C the deep lung air contains a concentration of alcohol proportional to the concentration of alcohol in the blood.
Average breath to blood ratio used.
2100 parts air. 1 part blood
Shallow blow
When a subject provides a breath sample that meets the minimum sample acceptance criteria, but is not a sample of deep lung air.
What is the margin of acceptability of Alcohol Standard results?
10% of the target value
EC means
IR means
Electrochemical
Infrared
Case fan
Cools the instrument
Pressure sensor tubing purpose
- Monitor the flow rate and volume.
2. Monitors ambient atmospheric pressure.
What radio frequency protection is in place
Faraday cage over circuit board
Metal case
RF detection
Sample acceptance criteria
- Minimum flow rate of 12L/min
- Minimum volume of 1.5 L
- 5% drop in flow rate after minimum volume is reached
Criteria for a Proper Breath Test
Two suitable breath samples within 20 MG% of each other.
Two diagnostic tests that have passed.
Two alcohol standard tests within 10% of the target value.
Alcohol standard simulator temperature.
33.8 C to 34.2 C
When is a quick test required?
After an equivocal refusal to confirm that instrument was working.
F2 function
5 logs
Personal log Instrument usage report Maintenance log Alcohol standard change log Breath test bulletin
Restart the observation period when:
The subject is not maintained in close proximity and in the field of view of the observer.
If the subject places anything in their mouth, burps, or vomits.