Ct Imaging Flashcards
are being used to create a gradient in the intensity of the X-ray beam
They are sometimes called “bow-tie” filters
They are mounted close to the X-ray tube.
The purpose of is to ⚫ reduce the dynamic range of the signal recorded by the CT
detector • Reduce the dose to the periphery of the patient
• Attempt to normalise the beam hardening of the beam - to aid with calibration
Beam shaping filters
CT scanner detectors
⚫___-__ detector elements along the detector arc •
800-1000
___-___detectors along z-axis
1-320
CT detectors are ___ in the axial plane (x-y plane), and ____ along the longitudinal axis (z-axis)
curved
rectangular
2 Detector type of CT & THEIR EFFICIENCY
Xenon gas filled-70%
Solid state- approaching 100%
Better detection efficiency
Solid state detector
Solid state detectors are generally
____
the photons interact with the detector and generate light.
scintillators
The light is converted into an electrical signal by ____
photodiodes
The septa and the strips of the anti scatter grid should be as __ as possible, since they reduce the effective area of the detector and thus reduce the detection of X-rays.
small
Detector sizes are the effective size at the ____
iso-centre
The minimum number of detector elements should be approximately ___to achieve a spatial resolution of d in the reconstructed image
2 FOV)/d
→___- detector elements are required to achieve aspatial resolution of 1 mm within a reconstructed image at a field of view of 400 mm
800
Spatial resolution can be improved by use of the ______ or_____
quarter detector shift
Dynamic focal spot
By shifting the detector elements by a distance of a quarter of the size of the detector elements, the theoretical achievable spatial
resolution becomes twice as good. It is generally implemented in detectors of all CT scanners.
Quarter detector shift
• Focal spot position on anode is rapidly oscillated during gantry rotation, doubling the number of projections
Dynamic or flying focal spot