CT Flashcards
What is partial volume effect?
What does it depend on?
What is the effect of partial voluming on low contrast objects?
A high contrast object that is smaller than the voxel may be seen larger
It depends on thickness of the slice,
Thinner slice provide better resolution
Partial volume effect may also reduce visibility of low contrast details.
Principle components of the scanners
- X-ray tube, collimator and detector are mounted in a rotating gantry.
- The rotation axis is the z axis.
- The x ray beam is collimated as a wide fan beam sufficient to cover the patient cross section at its widest
- It has a narrow width parallel to the z axis that for a single slice scanner defines the images slice thickness.
- With detector rays , the radius of the arc is equal to the focal distance. Total number of detectors 500-1000
What is the plane of the gantry rotation?
The plane of gantry rotation is perpendicular to long axis of tabletop.
It is possible to tilt the gantry by up to 30 degrees especially for head scans..
X-ray tube , position of the anodes cathode axis
Anode cathode axis parallel to the axis of rotation of scanner. This is to minimise the effect of the anode heel effect.
How many focal spot size? What is the smallest size?
What is the heat capacity?
Two focal spot size, the smallest being about 0.6mm
Heat capacities of 4MJ or more
What sort of filtration is used in CT?
Generally scanners have a total tube filtration of 6 mm aluminium
Earlier scanners used additional copper filters
Ideally a mono energetic beam would be used in CT.
CT matrix size
What is a CT number?
What is used for calibration of of CT number scale?
Variation in attenuation coefficient of each tissue relative to water is being dependent on what?
512x512 or 256x256 or 1024x1024
Average linear attenuation coefficient
Air and water
Variation in attenuation coefficient of each tissue relative to water being dependent on Kv and filtration of x ray beam.
Colimator
The beam is collimated to a fixed width ~50cm at the axis of rotation, sufficient to cover the full cross section of the patient
On z axis the size is variable and in the case of single slice scanner defines the image slice thickness
When would you use a post patient collimator?
Why would you use a post patient collimator?
For single slice scanners there might be a post patient collimator in front of the detectors.
Its purpose is to reduce scatter when the slice thickness is less than the width of detectors
Or to get accurate thickness selection for the smallest slice width.
Bow tie filters
Elliptical body shape
Low dose at the centre and unnecessarily high at the peripheries.
Thin at the centre and thicker at the periphery
Requirements for scanner detectors
- Small to allow good spatial resolution. For single slice scanners with 600-900 individual detectors are in detector bank, the width of each detector
What was the main detector type prior to development of the multi slice scanners ?
Ionisation chambers such as xenon or krypton because of high atomic number and k shell binding energy
Xe ( Z 54 and k shell 35 kev) at high atmospheric pressure 20atm
Detection efficiency 60%
Is ionisation chambers suitable for multi slice scanners?
What sort of detectors are used?
No
Solid state detectors
What is a solid state detectors?
They incorporate a scintillation with embedded silicon photodiode to detect the light output
They have negligible after glow ie signal lag and they have a stable response.
What is the scintillant is made up if ?
The scintillant is made up of
Cadmium tungstate
Bismuth germinate
Rare earth ceramics
What’s the detection efficiency of solid state detectors vs ionisation chambers?
Solid state detectors 98%. Although it is effectively less due to effectors have to be separated to prevent light cross over therefore the actual efficiency is closer to 80%
60% for ionisation chambers
The number of measurements taken in scanning a single section of the patient depend so what ?
Depends on :
Number of detectors
And number of measurements that are taken in the full rotation.
What is the most common method used for image reconstruction?
Filtered back projections
Is it possible to collect data over a reduced arc down to 180?
Give example of usage
Yes, this may be used for imaging rapidly changing scenes such as CT fluoroscopy and for cardiac imaging
What is a topogram/ scout view/ or scanogram?
This is a transmission image taken at a fixed projection angle.
The collimator is set to the narrowest slice width and the image is obtained as the patient table is moved through the gantry.
The resultant image has a poorer spatial resolution compared with the standard radiographs. The adv is minimal scatter.
Why is the image different on scout view to the conventional radiography?
Because there is no beam divergence along the z axis.
There is minimal scatter
These are used for planning the ct sequences including the selecting the start and end points and displaying scan slice position.
What is a CT fluoroscopy?
This is a display of a CT image in real time.
This is achieved by continuous rotation of the gantry without table movement.
Using fast reconstruction techniques from 180 data sets frame rates of 5 frames per sec or greater might be achieved.
Give example where CT fluoroscopy is used.
Biopsy needle placement
Although effective dose may be less than standard diagnostic scan, one should be aware of the high patient skin doses.
What is the other name of axial scanning?
Step and shoot
ie patient being moved along the z axis of the scanners between each succeeding slice
What does a slip ring technology entail?
It comprises a metal ring mounted on the gantry. This is connected to the signal output from the detectors. Adjacent to the gantry there is a connector that is able to retain good contact with the ring as the gantry rotates.
This connector takes the signal and passes it to the computer system for image reconstruction.
Slip ring technology
The detector signals have to be sent sequentially at high frequency in order to accommodate the total number of individual detectors on the gantry and the sampling frequency
Additional rings are used to connect the X-ray generator to the external voltage supply
What is the advantage of continuous rotation?
To allow faster rotation times.
To allow continuous acquisition of data while the table moves the patient through the scan plane.
Data for the complete volume can be collected in a single exposure.
What is the continuous acquisition of data called?
Volume , spiral or helical CT
What is a pitch?
A pitch is a table movement per rotation divided by slice thickness
T/ F?
The slice width can be less than the detector width.
False
The slice width cannot be less than the detectors width.
Advantages of helical scanning.
Scan speed- possible to do in single breath hold.
This reduces the risk of slice misregistration. Teh other adv are patient throughput and the reduced use of contrast medium.
What are the advantages of increasing pitch more than 1?
Reducing exposure time and reducing patient dose.
However the resolution is reduced because of the need of greater interpolation.
Pitch greater than 2 gives unacceptable image quality.
T or F
True 3D imaging requires isotropy that is the voxel size must be equal in all three dimensions.
True