CRT Flashcards
1
Q
AV delay optimization
A
2 methods :
- Ritter method :
AV OPT = AV short + d
Where d=(AVlong - AVshort) - (QApre - QAspont)
Set short AV , measure QApre (time from QRS onset to the premature end of A wave)
Then set long AV : QAspon (time from QRS onset to the end of A wave) - Ishikawa method :
Set long AV delay, measure the period between end of atrial emptying and beginning of aortic ejection, then substract it from the set AV interval