CPQR Flashcards
CPU/server test frequency
quarterly
digitizer test
2 mm tolerance, 3 mm fail
quaterly
plan transfer test frequency
quaterly
-to machine
-to other systems
plotter/printer test
Check the dimensions on the printout against the inputted contour and previous plot
-2mm/3mm
-more stringent if being used for patient setup
-quaterly
back-up recovery test
-check integrity of backed up files
-quaterly
CT geometry/density test
-2mm/0.2 and 3mm/0.3
-Check that the CT geometry and the relationship between CT number and density have not changed. Tolerances and action levels are specified in millimetres/relative electron density
annual TPS tests
revalidate data (check standar set of at least 4 plans)
independent quality control review
what does revalidating data consist of?
-check constancy of standard set of plans
-check repeatability of DVH
-check constancy of PDD with TMR data (TPS vs measured)
-E2E test