CoP for Non Standard linacs Flashcards

1
Q

What field do you need to create if your linac can’t make a 10x10 field?

A

A machine specific reference field

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What factors need to be corrected?

A

The quality index and field size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What are the reference factors for Tomotherapy?

A

5cm deep, 85cm SSD, phantom width>= field+5cm, backscatter>=5cm.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

If a machine is unable to make a large enough field for reference dosimetry what is an alternative?

A

Make a uniform field out of composite non-uniform fields. This is called a plan class specific reference

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What are the requirements of composite fields?

A

Must ensure the chamber reacts well to composite fields.

Needs to produce a uniform field large enough for reference dosimetry.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

For composite fields what term is added to the equation D=RN?

A

kfc - accounts for difference in chamber response between machine specific reference and plan class specific reference.