Computed Tomography Flashcards
This is the moving part of the scanner apparatus (CT).
Gantry
In the first years of CT, X-ray tube and detectors used linear scanning trajectory.
This data that were collected in this scanning pattern corresponds to attenuation measurements that are parallel to each other, and this geometry is called what?.
Parallel beam projection
This refers to the data collected at a specific angle of interrogation of the object, and this term is synonymous with the term profile or view.
Projection
Rays are individually attenuation measurements that corresponds to a line through the object defined at one end by the x-ray source and at the other end by a detector.
A projection is a collection of rays.
This refers to a fan of data that converges on a vertex , and using rotate-rotate geometry, the apex of the fan is the X-ray tube.
The individual rays corresponds to each detector measurement.
Fan beam projection
There are more detectors.
Minimal divergence of x-ray beam trajectory in and out of the plane figure.
What are the different beam geometries used in CT scan?
Parallel beam geometry
Fan beam geometry
Cone beam geometry
This is the center of the rotation gantry, and in most cases (but not all), it also the center of the reconstructed CT image -that is, pixel on the 512 x 512 reconstructed CT image.
Isocenter
Once the patient is on the table and the table is moved into the gantry technologist performs a preliminary scan called what?
CT radiograph.
This image is also called the scout view, topogram, scanogram, or localizer.
This is the basic step-and-shoot mode of a CT scanner.
The axial (also called sequential) CT scan
The gantry rotates at typical rotation speeds of 0.5 s or so, but the X-ray tube is not turned on all the time.
With the advent of MDCT, it is common to acquire contiguous CT images during axial acquisition.
With this kind of scanning, the table moves at a constant speed while the gantry rotates around the patient.
Helical (also called spiral) scanning
The advantage of helical scanning is speed - by eliminating the start/stop motion of the table as in axial CT, there are no inertial constraints to the procedure.
This describes the relative advancement of the CT table per rotation.
Pitch
For most CT scanning, the pitch can range between ___.
0.75 and 1.5
A pitch lower than 1.0 results in overscanning the patient and hence higher radiation dose to the patient than a pitch of 1.0, all other factors being equal.
A pitch greater than 1.0 represents underscanning, and results in lower radiation dose to the patient.
Pitch settings near _____ allow for faster scanning and are used for thoracic or paediatric CT scanning where speed is important.
1.5
____ pitch values are used in cardiac imaging, or when a very large patient is to be scanned and the other technique factors (kV and mAs) are already maximized.
High or low value?
Low pitch values
This is used to evaluate vascular perfusion and other physiological parameters related to blood flow to a specific organ.
CT perfusion imaging
This allows a CT scanner to repeatedly image a volume of tissue that is wider than the detector array.
Shuttle mode
Shuttle mode imaging is a hybrid between CT perfusion and helical CT imaging - the table shuffles back and forth over a relatively small (100 to 200 mm) predefined region in the patient, allowing temporal imaging and hence CT perfusion assessment to be made over a wider length of anatomy than the width of a MDCT.