Self inflating resuscitation device indication
PPV
Self inflating resuscitation device can be used with or without ETT
Yes. Self inflating resuscitation device can be used with without ETT.
Adverse effects of self inflating resuscitation device
Pneumothorax
Facial trauma
Gastric inflation
Parts to self inflating resuscitation device
Gas inlet
Gas outlet
Gas tubing
PEEP VALVE
Manometer
Pressure release valve
Oxygen reservoir
Valve assembly
Advantages of her tea piece flu inflating
Does not need gas source
Disadvantages of a teepee slow inflating
Cannot be used to provide free flow/ CPAP
Role of the oxygen reservoir in the self inflating device
Can provide higher FiO2
How does the self inflating device work
When squeeze it directs gaster the valve assembly to the patient
Re- expands
From room air
Gas tubing
Oxygen reservoir
Desired sizes for the self inflating device
neonates
pediatric
450
1000
What is the flow of the self inflating reservation device
10 L per minute
Maximum pressure relief control and neonate
Full-term 40 cm of water
Preterm 30 cm of water
What is the desired peep for the self inflating bag
5 cm of water
What should you do before using on newborn?
Ensure proper settings:
Pippa 5 cm of water
Flow rate 10 L per minute
Proper size bag
Proper size mask
Maximum pressure relief control
Flowing inflating resuscitation device
Also called?
Anesthesia bag
Non-self inflating
Indications for the flow inflating resuscitation device
PPV
CPAP/PEEP
Free follow supplemental oxygen