Cochlear implants Flashcards

1
Q

Team of developers

A

engineers, otologists, audiologists, psychoacousticians, and neurophsysicits

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

how does it work?

A

Cochlear implants bypass damaged peripheral hearing system and directly stimulate the CNVIII

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

indicaters a mapping appointment is needed

A
changes in auditory discrim
increased repetition needed
adition/omission of syllables 
prolongation of vowels
change in vocal quality
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

telemetry

A

confirms proper communication of processor and electrodes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

neural response assessment

A

electrophysiologic response from the nerve in response to electrode stim
especially useful in children

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

t levels

A

measure of threshold for sound on each electrode (cochlear)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

C or M levels

A

comfort levels using a scale

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

speech strategies

A

different methods of stimulation can produce different perceptions from the patient
ex simultaneous sequential or both

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

condition

A

AB only - sends stim to all cannels at the same time. it can push away any buildup around array
activation?

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

impedance

A

opposition to electrical current flow across an electrode
voltage/current measured in kOhms

influenced by electrode and lead, but also the medium
fibrous tissue, electrolytes, macrophages, proteins

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

voltage

A

current x impedance

current stays constant, increases in impedance will cause an increase in voltage.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

high impedance

A

open
may be due to air bubbles, send stim then measure again?
hopfully its temporary

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

short circuit

A

low impedance, usually a phsyical breakdown of two channels permanently
usually only occurs at implantation
compare over time, note dips in impedance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

NRI (ab)

A

corresponds well to M levels

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

NRT (cochlear)

A

not always correlated to C levels, can sometimes be more in the middle

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Uses of NRT/NRI

A

relatively stable over time

used with impedances to tell if change in performance is due to device function or neural responsiveness

get within first few months of stimulation to use as baseline

assess for internal device failure, if you had them before and nwo you dont it could be a soft failure
can also use this so determine what pitch abilities.

highly influences by neural survival, synchronous neural activity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

threshold levels - (cochlear)

A

at or jsut above threshold, ascending bracketing loudness growth chart

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

med el threshold levels

A

highest stim where no sound is perceived

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

sound detection thresholds

A

15-20 for cochlear

30-40 others

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

c levels

A

set below maximum confort level (summation)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

m levels - med el

A

highest stim level at which sound is loud but comfortable

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

m level - AB

A

most comfortable level

soundwave 2.0 uses speech bursts for m levels

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

methods of loudness balancing

A

balancing and sweeping
may say “sounds good but something is weird”
doesnt always do this at activation,
usually at about 80% of MCL? at or near upper stim level

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

sweeping

A

start at one end and start just below of every channel as itmoves through the array going low to high, “pitch should get higher, loudness should not.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
pitch ranking
if the electrode array is rolled over on itself, pitch does not increase from low to high on sequential electrodes reorder channels if necessary to match the perception
26
compliance
the amount of voltage allowed for each electrode
27
"out of compliance"
max voltage available from the implant is not suffiecient to generate the desired current level insuffient loudness growth, variable loudness, sound may be distorted, poor battery life, decrease performance in general dont want your c level to be just below compliance if you cant you may need to increase pulse width some days it sounds good and some days not.
28
automatic pulse width (APW) - ab
calculates and adjusts pulse width and rate pased on compliance and M-level requirements it can do this when the patient is not in front of you so if you have fluctuating impedance you get more compliance headroom so you may want to turn on
29
when to use manual pulse width?
poor sound quality cannot obtain adequate loudness due to facial nerve stim fluttering or choppy
30
maxima
takes incoming signal to spectogram and pulls our where the most energy is, usually 8-10, transmitted via T levels of the map
31
loudness growth - amp
as long as there is sufficient voltage available, current amp can be increased and the precept of loudness growth will result if insufficient voltage is not available to meet the current requirement, the channel will be out of compliance. (no further loudness growth)
32
loudness growth
amplitude, pulse width, rate
33
loudness growth - PW
short (narrow) pulse widths produce higher thresholds and steeper loudness growth curves. longer pulse durations (wider) produce lower thresholds and larger dynamic ranges, due to slower loudness growth near threshold.
34
loudness growth - rate
pitch changes due to rate saturate at 500 PPS | higher stim rates require narrow PW which consumes sig power.
35
eCAP
electrically evoked compound action potential potential that reflects fynchronious firing of large # of electrically stimulated nVIII fibers started by putting needle electrode, cochlear discovvered with Nucleus to use intracochlear to stimualte instead N1 ~.2-.5 ms followed by P1 ~ 1 ms amp 20-1500 uV
36
utility of eCAP
integirty of system, how are the devices talking and how is the nerve responding. at activation, once you get all channels you dont really have to do it that much, more useful in children series of sounds, get louder and try to tolerate it these are the highest levels so if you dont get facial nerve stim that you probably wont, also can start to get pitch information
37
NRT/NRI high influenced by
neural survival synchronous neural activity nerve survival with respect to the measuring electrode.
38
eSRT
electrically evoked stapedial reflex threshold SRT occurs at or near the M/C levels levels used by the processor bilateral contraction so stim the implant and measure non implanted ear need to consider middle ear history especially for bilateral
39
how to administer eSRT
probe in contra ear, continue to record acoustic admittance with 226 probe tone, present programming stimulus for upper limit of DR change in admittance occurs time=locked with stimulus when presentation level is of ideal intensity of C/M level measure as many as you can, then interpolate reduce c/m for live mode then increase to eSRT level
40
speech encoding strategy
method which the implant translates the incoming acoustic signal into patterns of electrical pulses provide spectral and envelope info
41
stimulation mode
location of the reference electrode to the active electrode -monopolar vs bipolar med el monopolar only
42
threshold levels
ensure soft sounds are audible sound field responses are in line with specs perception of soft softs if soft sounds are too loud bring down T
43
T-Tail
limited change in loudness perception across a wide range of stimulation ltevels in the lower part of the EDR (electrical dynamic range)
44
T levels for tinnitus
Set t levels over level of tinnitus, then reduce globally, loudness balancing at 50% over t levels
45
interpolation
estimation of stimulation levels on the basis of measure values from neighboring channels.
46
programming method
behavioral, objective preset (using NRT or NRT to set up map) objective w/ behavioral offset help set configuration measure on T then interpolate
47
influences on levels
speech processing strategy bipolar/monopolar stim rate C and M levels (current) proximity of electrode array to modiolus
48
non auditory stimulation?
current can reach neighboring nerves, run through measurements and summate to make sure, test loudness clapping and jingling keys reduce levels or deactivate electrode if causing coughing or favial tiwtch. if you are way out of your DR then just turn off, otherwise drop 5 CU
49
the need for power
``` sufficiant voltage available to the implant to ensure that all electrodes are in compliance sufficient power (voltage and current) for the demands of he chosen MAP ```
50
power optimization
can the battery provide sifficient voltage to deliver the requested amount of current calculates amount of power needed in worst case condition (ambient noise >65) for all electrodes to be within compliance
51
problems with battery life
increase pw | decrease rate/maxima if high stim levels or compliance conditions are not an issue
52
live voice modifications -----
tilting decrease or increase levels on all channels gain frequency adjustments
53
tilting
increase high 2 and decrease low 2 rather than decreasing low 4 steps
54
hearing low level humming?
take down the c levels for the lows, but they are still hearing if then it could be their own voice so bring down T levels.
55
sensitivity
distance of hearing, perception in noise
56
processing strategy
set of rules and processes for converting the acoustic input signal into electrical stimulation waveforms paired or sequential
57
paired or sequential
paired is faster so you may do sequential for older patients
58
intensity domain
``` influenced by how sound is captured and processed directional mics mic placement AGC input dynamic range ```
59
spectral domain
nromal hearing 20-20kHz, implants 250-8000 Hz music is different, notes that are close in frequency can processed through the same ci filter bank so perception is the same, larger difference to differentiate
60
the 5 processing strategies
``` continuous interleaved sampling (CIS) HiResolution (HiRes) - AB Advances combination encoder (ACE) - cochlear Spectral peak (SPEAK) Fine structure processing (FSP) -mel el ```
61
CIS
filters sounds using band pass filters envelope compressed non linearly 1000 pps pulses sent sequentially
62
HiRes
high stim rate and up to 16 channels up to 5000 pps uses half-wave rectifier for envelope detection current steering, using multiple electrodes in different patterns for different frequencies, more sophisticated AGC algorithm new versions - increase virtual channels improve spectral res and speech recognition in noise, soud quality and music appreciation designed to reduce battery consumption
63
n of m
acoustic energy in each m channel is determined and stimulation is administered to only the n channels with the highest amplitude inputs. reduces number of active electrodes = faster stim rates, increase batt life
64
SPEAK ACE
channel selection scheme envelope signals for different channels re scanned prior to each from of stiulation, the highet amplitudes are identified and stimulus is delievered to those electrodes. fixed parameter ACE 1000 pps and SPEAK 250 pps
65
FSP
new form of CIS offered by med el intermetiate pitches generate via bell shaped overlapping filters diesigned to improve perception of high frequency phonemes based on place cues
66
selecting speech processor strategy
high rate = ace FSP HiRes sig research indicates that performance is improved with high rate strategies preference is strongly correlated with performance
67
voltage compliance
each system has a finite voltage capacity (size and type of battery) which determines the maximum amount of current it can deliver. The amount of power available to operate the implant and the current that will be delivered to an individual electrode is determined by the efficiency of the signal delivery from the external coil to the internal device, the voltage capacity of the battery, and the electrode impedance. The maximum amount of electrical current available to stimulate an electrode is determined by Ohm's law The max amount of current that can be delivered to an electrode by given power source. The voltage is determined by capacity of the battery so its fixed. the current available at each electrode is variable across recipients and electrodes by the impedance of each electrode high impedance will result in lower amount of current available to stimulate. lower electrode impedance results in a higher amount of current available o stimulate the electrode