cobas4000 Flashcards

1
Q

What are the elements of the cobas4000 system?

A
c311 clinical chemistry module
e411 immunoassay module, which comes in a rack or disk version
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Is the cobas4000 system connected or modular?

A

The two systems are not connected together.

The c311 stays where it is installed.

The e411 disk system stays where it is installed.

The e411 rack system can be benchtop.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What size of customer is the cobas4000 appropriate for, and what volumes?

A

Low volume customer

<500,000 tests per year
or
50-200samples per day

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How could the cobas4000 system be used?

A

Primary system for physician office labs or low volume testing needs

Backup system for the cobas6000 system

Niche place,ent for STAT or specialized testing:

c311: HbA1c, DAT, TDM
e411: cardiac, fertility, ioPTH

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What reagent packs are used for the cobas4000 system?

A

The c311 uses the cobas cpacks
50-800 tests per pack
98% Ready to use
Onboard stability of 67 days

The e411 uses the cobas epacks
100 or 200 tests per pack
Ready to use
Onboard stability of 47 days

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Describe cobas epacks

A

100 or 200 packs

Onboard stability of 47 days

Calibration frequency once per lot

Ready to use?

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Describe cobas cpacks

A

50-800 tests per pack

Onboard stability of 67 days

Ready to use

Calibration frequency of once per lot

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

How many reagent channels are on the cobas c311?

A

42 reagent channels and 3 ISEs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

How many reagent channels are on the e411?

A

18 reagent channels

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What are the areas of the cobas c311.

A

Sample area
- sample probe has abnormal descent, clot detection and liquid level detection

Reaction area
- ultrasonic mixing minimizes cross contamination

Reagent compartment

Control unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What are the components of the e411 system?

A

Sample area
- liquid level and clot detection

Reagent compartment

Reaction area

  • ECL technology
  • disposable tips and cups minimize carryover

Control unit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What are the TATs for the cobas4000 system?

A

3-10 minute durations on the clinical chemistry side

Immunoassay side:
Most are finished in 18 minutes, with a handful of 27 minute assays. Additionally we have 7 STAT assays that are finished in 9 minutes, including STAT PTH, STAT HCG and STAT Cardiac

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Explain ECL technology

A

This proprietary method to Roche is a way of keeping the light on longer throughout the reaction. This allows for a more stable and wide measuring range, increased sensitivity due to the constant signal, and accurate detection of the targeted antibodies or antigens being measured.

This technology also allows for a smaller sample size which leads to fewer redraws and increased patient convenience.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What are the three ISEs?

A

Sodium
Potassium
Chloride

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What is cobas link?

A

This is a way for support to dial in and assist with troubleshooting

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is the sample volume requirements for the cobas4000?

A

10-50 micro liters

17
Q

Dimensions of e411 rack

A

W 67”
D 37.4”
H 43”

Weight: 551lbs

18
Q

Dimensions for e411 disk

A

W 47.2”
D 28.7”
H 43”

Weight: 397lbs

19
Q

Throughput of c311

A

300 photometric tests per hour

450 ISEs per hour

20
Q

Dimensions for c311

A

W 52”
D 34”
H 67”

Weight: 550lbs

21
Q

Throughput of e411

A

Up to 86 tests per hour

22
Q

How many samples can be loaded at once on the cobas c311.

A

110 at once
User defined STAT positions

Accepts primary tubes as well as 1.5ml microcups

23
Q

Explain the reaction wheel on cobas c311

A

There are 66 disposable cuvettes which are immersed in a 37° bath. Sonic wash is used to reduce chance of carryover

24
Q

How many patient samples can be loaded onto cobas e411?

A

Rack system allows up to 75 patient samples at one time using the hitachi 5-position racks

Disk system allows 30 patient samples on board at one time

STAT port available on both and accept primary tubes and 2ml hitachi cups

25
Q

Describe tips and cups on cobas e411

A

You can load up to 350 disposable tips and 180 disposable cups