Cleaning and Decontamination Flashcards
What is the preferred pH for detergents used for most cleaning processes?
Neutral pH
Which of the following statements about the use of mechanical washers is NOT true?
All items should be washed on the same cycle
Water must be degassed each time the ultrasonic cleaner’s tank is changed, because
excess bubbles from filling reduce the energy of the cavitation process
Written celaning instructions for surgical insturments should be provided by
the device manufacturer
Pyrogens
are fever-producing substances
Untreated water
increases the likelihood of mineral scale deposits
These are used to breakdown fatty tissue on instruments.
Lipase enzymes
Instrument lubrication is performed
after cleaning
To prevent coagulation of proteins, instruments should be pre-rinsed using
cool water
Detergents used in mechanical cleaners should be
low foaming
Instuments should be cleaned using a
to and fro motion
Powered surgical instruments
cannot be immersed
The decontamination area should have
negative air slow in relation to the other areas of the department
The temperature in the decontamination are should be between
60 to 65 degrees fahrenheit
Rigid contaner filter retnention plates should be
removed from the contaner and lid and cleaned separately