Chapter 6: Sampling, Standardization, and Calibration Flashcards

1
Q

are traditionally classified as gravimetric methods, volumetric methods, or instrumental methods

A

quantitative methods

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

an analysis used for samples whose masses are greater than 0.1 g

A

macro analysis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

is performed on samples in the range of
0.01 to 0.1 g

A

semi-micro analysis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

is performed on samples in the range of
10^-4 to 10^-2 g

A

micro analysis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

is performed on samples in less than 10^-4 g

A

ultramicro analysis

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

TRUE or FALSE
Techniques for handling very small samples are quite different from those for treating macro samples.

A

TRUE

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

species present in the range of 0.01 to 1% are usually termed

A

minor constituents

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

species present in the
range of 100 ppm (0.01%) and 1 ppb are called

A

trace constituents

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Components present in amounts lower than 1 ppb are usually considered to be

A

ultratrace constituents

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

The analysis of real samples is complicated by the presence of the

A

sample matrix

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

can contain species with chemical properties similar to the analyte. can react with the same reagents as the analyte, or they can cause an instrument response that is not easily distinguished from the analyte.

A

Matrix

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

If the interferences are caused by
extraneous species in the matrix, they are often called

A

matrix effects

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

The process of acquiring a
representative fraction is termed

A

sampling

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

is often the most difficult
aspect of an analysis.

A

sampling

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

TRUE or FALSE
The composition of the gross
sample and the laboratory
sample must closely resemble
the average composition of
the total mass of material to be
analyzed

A

TRUE

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Steps in obtaining a lab sample

A

identify the population
collect a gross sample
reduce the gross sample to a laboratory sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

representative portion of a whole analytical sample, which with further treatment, becomes the laboratory sample

A

gross sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

The items chosen for analysis are often called

A

sampling units or sampling increments

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

chemists usually call
the collection of sampling units or increments the

A

gross sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

gross sample is usually reduced in size and
homogenized to create the

A

laboratory sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Statistically, the goals of sampling process are

A

obtain a mean analyte concentration for an unbiased estimate of population mean
obtain a variance in the measured analyte concentration for an unbiased estimate of the population variance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

can reveal whether the between
samples variation (sampling variance plus measurement variance) is significantly
greater than the within samples variation (measurement variance).

A

analysis of variance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

TRUE or FALSE
Errors due to invalid sampling
are unique in the sense that they are not controllable by the use of blanks and
standards or by closer control of experimental variables. For this reason, sampling errors are usually treated separately from the other uncertainties associated with an analysis

A

TRUE

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

is a miniature replica of the entire mass of material to be analyzed. It should correspond to the bulk material in chemical composition and in particle-size distribution if the sample is composed of particles.

A

gross sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
gross sample is acquired by
1. uncertainty that can be tolerated 2. degree of heterogeneity of the whole 3. level of particle size, which heterogeneity begins
26
is a state intermediate between a suspension and a solution.
colloid
27
intermediate between heterogeneity that develops in particles that may have dimensions on the order of a centimeter or more and maybe several in grams are called
colloidal materials and solidified metals
28
the constant Ks is called the
Ingamells sampling constant
29
is a container that can be opened and filled at any desired location in the solution.
sample thief
30
TRUE or FALSE With large volumes of solutions, mixing may be impossible. It is then best to sample several portions of the container with a “sample thief,”
TRUE
31
What is used to collect vapor samples
gas-tight syringe or stainless steel canister
32
TRUE or FALSE The laboratory sample should have the same number of particles as the gross sample.
TRUE
33
samples of metals are obtianed by
sawing milling drilling
34
TRUE or FALSE With some materials, a representative sample can be obtained by sawing across the piece at random intervals and collecting the “sawdust” as the sample.
TRUE
35
the steps in sampling a particulate solid includes
crushing grinding seiving mixing dividing the sample to reduce its mass
36
can lead to higher throughput (more analyses per unit time), higher reliability, and lower costs than manual sample handling.
automated sample handling
37
two different methods for automated sample handling
batch or discrete approach continuous flow approach
38
A method where Automated instruments that process samples in a discrete manner often mimic the operations that would be performed manually.
Discrete methods
39
In continuous flow methods, the sample is inserted into a flowing stream where a number of operations can be performed prior to transporting it to a flow-through detector. Hence, these systems behave as automated analyzers in that they can perform not only sample-processing operations but also the final measurement step.
continuous flow methods
40
discrete methods do
dilute filter partition grind centrifuge homogenize extract treat samples with reagent
41
continuous flow method do
reagent addition dilution incubation mixing dialysis extraction
42
2 different types of continuous flow systems
segmented-flow analyzers flow injection analyzers
43
divides the sample into discrete segments separated by gas bubbles
segmented-flow analyzer
44
is a more recent development, samples are injected from a sample loop into a flowing stream containing one or more reagents
flow injection analysis
45
is a band-spreading or mixing phenomenon that results from the coupling of fluid flow with molecular diffusion.
dispersion
46
is mass transport due to a concentration gradient.
diffusion
47
the sample and reagent are both injected into carrier streams and merged at a tee mixer.
merging zones flow injection analysis
48
sample dispersion is controlled by the sample size, the flow rate, and the length and diameter of the tubing. It is also possible to stop the flow when the sample reaches the detector to allow concentrationtime profiles to be measured for kinetic methods
flow injection analysis/ merging zone FIA
49
determines the relationship between the analytical response and the analyte concentration
calibration
50
the relationship of analytical response and the analyte concentration (calibration) is determined by
chemical standards
51
Constituents of a sample other than the analyte are called
concomitants
52
Concomitants that interfere with the determination of an analyte may be referred to as
interferences or interferents
53
needs to indicate whether the threshold has been exceeded
comparator
54
Some analytical procedures involve comparing a property of the analyte (or the product of a reaction with the analyte) with standards such that the property being tested matches or nearly matches that of the standard are called as
null comparison or isomation methods
55
are among the most accurate of all analytical procedures, the analyte reacts with a standardized reagent (the titrant) in a known stoichiometric manner.
titration
56
a type of calibration process in which a series of standard solutions is prepared separately from the sample, The standards are used to establish the instrument calibration function, which is obtained from analysis of the instrumatrment response as a function of the known analyte concentration.
external standard calibration
57
A plot of instrument response versus known analyte concentrations is used to produce a
calibration curve or working curve
58
provides the means for objectively determining such a line and also for specifying the uncertainties associated with its subsequent use
regression analysis
59
Least squares method two assumptions
linear relationship between the measured response y (absorbance in Figure 6-10) and the standard analyte concentration x any deviation of the individual points from the straight line results from error in the individual measurements
60
The vertical deviation of each point from the straight line is called a
residual
61
simple least-squares analysis may not be appropriate when the uncertainties in the y values vary significantly with x. In that instance, it may be necessary to apply different weighting factors to the points and perform a
weighted least square analysis
62
is a rough measure of the magnitude of a typical deviation from the regression line.
SD abour regression/ standard error of the estimate/ standard error
63
measures the fraction of the observed variation in y that is explained by the linear relationship and
coefficient of determination R^2
64
The process of relating multiple instrument responses to an analyte or a mixture of analytes is known as
multivariate calibration
65
The least-squares procedure just described is an example of a
univariate calibration
66
The use of multivariate statistical methods for quantitative analysis is part of the subdiscipline of chemistry called
chemometrics
67
involves adding the interfering species to all the samples, standards, and blanks so that the interference effect becomes independent of the original concentration of the interfering species in the sample
saturation method
68
is a species, not itself an interfering species, added to samples, standards, and blanks in sufficient amounts to make the analytical response independent of the concentration of the interfering species
matrix modifier
69
is added that reacts selectively with the interfering species to form a complex that does not interfere
masking agent
70
can sometimes be used if the interfering species produces no significant effect below a certain concentration level.
dilution method
71
attempts to duplicate the sample matrix by adding the major matrix constituents to the standard and blank solutions
matrix-matching method
72
a type of calibration method in which a known amount of a reference species is added to all the samples, standards, and blanks
internal standard method
73
We use the _______________ when it is difficult or impossible to duplicate the sample matrix. a known amount of a standard solution of analyte is added to one portion of the sample
method of standard additions
74
additions of known amounts of standard analyte solution are made to several portions of the sample, and a multiple additions calibration curve is obtained.
multiple additions method
75
is a general term that applies to many fields including analytical chemistry. It is a numerical expression representing the performance or efficiency of a given device, material, method, or procedure.
figures of merit
76
Analytical procedures are characterized by a number of figures of merit
accuracy precision sensitivity detection limit dynamic range
77
the change in the measured quantity per unit change in analyte concentration. slope of the calibration curve
calibration sensitivity
78
Scatter in the response is also called as
noise
79
is the smallest concentration that can be reported with a certain level of confidence
detection limit
80
TRUE or FALSE A k value of 2 corresponds to a confidence level of 92.1%, while a k value of 3 corresponds to a 98.3% confidence level
TRUE
81
of an analytical method most often refers to the concentration range over which the analyte can be determined using a linear calibration curve
linear dynamic range
82
is a sequentialplot of some characteristic that is a criterion of quality.
control chart
83
if the value is within the upper control limit and lower control limit, the balance is said to be in
statistical control
84
determines the suitability of an analysis for providing the sought-for information and can apply to samples, to meth odologies, and to data
validation
85
is the final step before release of the results
data validation
86
a known amount of a standard solution of analyte is added to one portion of the sample
method of standard additions/ standard addition method
87
This standard additions method assumes a linear response
multiple additions method
88
is a rough measure of the magnitude of a typical deviation from the regression line
standard deviation about regression, also called the standard error of the estimate or just the standard error
89
is identical to the sample but without the analyte.
ideal blank
90
Most often a real blank is either a ___________, containing the same solvent in which the sample is dissolved
solvent blank
91
containing the solvent plus all the reagents used in sample preparation.
reagent blank
92
contains the solvent or the solvent plus all the reagents used to prepare the samples.
blank
93
is the ratio of the calibration curve slope to the standard deviation of the analytical signal at a given analyte concentration.
analytical sensitivity
94