Chapter 5 p. 2 Flashcards
SSFP stands for
steady state free precession
the SSFP method overcomes the problem of __ (__) in GE sequences and allows one to obtain __
inhomogeneity artifacts (T2* weighting); true T2 images
in SSFP, signals are digitized only from the __, not from the __; this is down by repositioning the __ so that it does not occur at the same time as the __
stimulated acho; FID; following excitation pulse
repositioning is achieved by applying a __, which speeds up __ so that the __ occurs sooner
rewinding gradient; rephrasing; stimulated echo
SSFP: the resulting signal has more __ than regular GE
T2 weighting
there are 2 Te values
the actual TE (from echo to next RF); effective TE (from echo to RF pulse which created the FID)
TEeff = __
2TR - TE
image from slide 35
the longer the __ TE the lower the __ TE
actual; effective
advantages of SSFP (2)
signal rephasing is initiated by an RF pulse rather than by a gradient, so the images have more true T2 weighting than conventional GE sequences; scans can be 2D or 3D
disadvantages of SSFP (3)
susceptibility to artifacts; image quality may be poor; loud gradient noise due to very frequent gradient application
images from slide 37 (not coherent or spoiled images)
coherent GE sequences sample __ to produce T1 or T2 weighting depending on __
both the FID and the stimulated echo; the TE
incoherent (spoiled) GE sequences sample __ to produce mainly __ images
the FID only; T1-weighted
SSFP sequences sample __ which is done before the __ to produce images which are more __
only the stimulated echo; FID; T2-weighted
image from slide 38 (same as slide 37)
in balanced GE, a sequence of gradients with __ is used to correct for __ in __ (2)
opposite polarities; phase inhomogeneities; CSF or flowing blood
moving H nuclei accumulate a __ as they pass along the gradients, which means that nuclei in __ (2) are __ and thus achieve high __
zero phase change; CSF and blood; coherent; signal intensity
balanced GE: the gradients are applied in the __ directions
slice and frequency encoding
balanced GE: larger __ and shorter __ are used compared to in coherent GE, resulting in higher __ and shorter __
flip angles; TR values; SNR; scan times
image from slide 39
balanced GE: to avoid saturation occurring due to __ (which increases __ contrast), the excitation phase pulse is __
large flip angles; T1; changed every TR
avoiding saturation is achieved by using a flip angle of __ in the first TR and then a __ pulse in the succeeding TR
45 degrees; 90-degree pulse
balanced GE: alternating pulses allows the transverse magnetization to __ after each repetition, and thus __ is avoided
have opposite phases; saturation
balanced GE, avoiding saturation: tissues with __ will be bright, resulting in high __, good __, and short __
T1/T2 ratios approaching unity; SNR; CNR between fat, water, and surrounding tissues; scan times
image from slide 40
balance GE is used for __ (3)
spine imaging; imaging of the auditory canal, and imaging of structures where CSF flow is present
balanced GE is also used to image __
blood vessels in the brain, heart, etc.
balanced GE requires a short __ to reduce __ and __ as well as a long __ to enhance __
TR; scan time and flow artifact; TE; T2* weighting
fast GE sequences are either __ or __, where the __ is substantially shorter than in conventional ones
coherent or spoiled GE sequences; echo time
FGE: only __ is applied, because it takes less time to __, and after this is done only __ (as in __)
part of the RF excitation pulse; turn on and off an RF pulse; part of the echo is read (as in partial echo imaging)
FGE: because only part of the RF pulse is applied during each repetition, __ are applied __ of the sequence in order to __ (__0
some extra pulses; before the beginning; pre-magnetize the tissue (load it with RF energy)
FGE: there are several ways to pre-magnetize the sample (2):
an 180-degree pulse is applied to completely saturate the sample; driven equilibrium approach
driven equilibrium approach
a 90/180/90 degree combination is applied before the pulse sequence