Chapter 5 & 6 Flashcards
Spatial encoding & k-space
The phase ______ determines the number of lines that must be filled to complete the scan
a. cube
b. matrix
c. parameter
d. encoder
b. matrix
The frequency encoding gradient is switched on when the signal is received and is also called the __________ gradient.
a. reading
b. readout
c. FOV
d. Uneven
b. readout
Regarding coronal or axial imaging of the head, the ___ gradient performs phase encoding.
a. Y
b. X
c. K
d. Readout
b. X
A generation of phase differences along a particular direction of a slice is:
a. Frequency encoding
b. Fourier transform
c. Phase encoding
d. Slice selection
c. Phase encoding
A greater number of phase encoding steps results in:
a. Reduced spatial resolution
b. Improved contrast resolution
c. More pixels in the FOV along the phase axis
d. Reduced number of pixels along the phase axis
c. More pixels in the FOV along the phase axis
The amplitude of the phase encoding determines:
a. Size of the FOV along the phase axis
b. The central portion of K-Space
c. The outer portion of K-Space
d. FOV along the frequency axis
a. Size of the FOV along the phase axis
Slice thickness is regulated by:
a. The transmit bandwidth
b. The receive bandwidth
c. Slice select slope
d. Both A & C the transmit bandwidth and slice select slope
d. Both A & C the transmit bandwidth and slice select slope
If the slice select gradient is turned on in the Y-Axis:
a. Coronal images are seen
b. Phase encoding gradient is along the X-axis
c. Precessional frequency increases
d. Both A & B
d. Both A & B coronal images are seen and phase encoding gradient is along X-axis
The phase encoding gradient in a spin echo sequence is turned on:
a. Twice per TR
b. Just prior to the rephasing pulse
c. At the same time as the echo peak
d. At the same time as the excitation pulse
b. Just prior to the rephasing pulse
Which of the following is not a function of the slice select gradient:
a. Slice position
b. Slice resolution
c. Slice orientation
d. Slice thickness
b. Slice resolution
In order to produce an oblique image:
a. The phase encoding gradient is steep
b. The frequency encoding gradient is shallow
c. The bandwidth is increased
d. Two slice select gradients are used
d. Two slice select gradients are used
The magnetic field gradient that turns on prior to the rephasing of the echo is the:
a. Slice select gradient
b. Phase encoding gradient
c. Rephasing gradient
d. Frequency encoding gradient
b. Phase encoding gradient
The readout gradient is responsible for:
a. The size of the FOV
b. The slice select slope
c. The phase encoding shift
d. The image resolution
a. The size of the FOV
Decreasing the slope of the slice selection gradient will have what effect on the resulting slice:
a. It will change the slice thickness
b. It will change the orientation of the slice
c. It will invert the phase and frequency directions
d. None of these
a. It will change the slice thickness
In order to generate an MR image, the slice selection gradient must be turned on during which of the following events:
a. Patient positioning
b. T1 relaxation
c. Collecting the echo
d. Application of the RF pulses
d. Application of the RF pulses
Which would be used for slice selection in order to produce a coronal head image?
a. X-gradient
b. Y-gradient
c. Z-gradient
d. A 90 degree RF pulse
b. Y-gradient
The applied magnetic field gradient activated during the initial RF excitation pulse is the:
a. Slice selection
b. Pixel selection
c. Frequency selection
d. Phase encoding
a. Slice selection
Increasing the slope of the phase encoding gradient produces:
a. Greater frequency
b. Large phase shift
c. Small phase shift
d. Large receive bandwidth
b. Large phase shift
Very steep phase encoding slopes ______ the amplitude of the resultant echo.
a. Increase by a factor of three
b. reduce
c. slightly increase
d. have no effect on
b. reduce
Gradients alter:
a. the magnetic field strength
b. precessional frequency
c. phase of nuclei
d. all of the above
d. all of the above